R&S NGM202 User Manual page 126

Hide thumbs Also See for NGM202:
Table of Contents

Advertisement

®
R&S
NGL200/NGM200
[SOURce:]VOLTage:PROTection[:STATe]
[SOURce:]VOLTage:PROTection:CLEar
[SOURce:]VOLTage:PROTection:LEVel
[SOURce:]VOLTage:PROTection:TRIPped?
[SOURce:]VOLTage:PROTection[:STATe] <state>
[SOURce:]VOLTage:PROTection[:STATe]?
Sets or queries the OVP state of the previous selected channel.
Parameters:
<state>
Example:
[SOURce:]VOLTage:PROTection:CLEar
Resets the OVP state of the selected channel. If an OVP event has occurred before,
the reset also erases the message on the display.
Example:
Usage:
[SOURce:]VOLTage:PROTection:LEVel <voltage>
[SOURce:]VOLTage:PROTection:LEVel?
Sets or queries the overvoltage protection value of the selected channel.
Parameters:
<voltage>
Example:
User Manual 1178.8736.02 ─ 06
......................................................................... 123
........................................................................... 123
............................................................................ 123
......................................................................124
OFF | ON | 0 | 1
OFF | 0
OPP is deactivated
ON | 1
OPP is activated
See
Example "Configuring the overvoltage protection"
on page 122.
See
Example "Configuring the overvoltage protection"
on page 122.
Event
numeric | MIN | MAX
numeric
Numeric value for the overvoltage protection value in V.
MIN
Minimum value for the overvoltage protection value at 0.00 V.
MAX
Maximum value for the overvoltage protection value at 20.05 V.
Range:
0.000 to 20.05
*RST:
20.05
Default unit: V
See
Example "Configuring the overvoltage protection"
on page 122.
Remote Control Commands
Configuration Commands
123

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ngm200 seriesNgl200 seriesNgl-com2aNgl201Ngl202Ngm201

Table of Contents