// Query the range of the overcurrent protection delay time
// ************************************************
CURR:PROT:DEL:INIT? MIN
// response: 10
CURR:PROT:DEL:INIT? MAX
// queries the upper and lower limit of the
// overcurrent protection delay time in ms
// response: 60000
// ************************************************
// Query a tripped overcurrent protection
// ************************************************
INST OUT1
CURR:PROT:TRIP?
//queries whether the OCP in channel 1 has tripped
//response: 1 OCP is tripped
//response: 0 OCP is not tripped
CURR:PROT:CLEar
//resets a tripped OCP in the selected channel
// ************************************************
// Link the electronic overcurrent protection of the channels logically
// ************************************************
INST OUT1
CURR:PROT:LINK 2
// links the overcurrent protection of channel 1 with channel 2
CURR:PROT:LINK?2
// queries the combined overcurrent protection of the selected channel
// ************************************************
// Unlink linked overcurrent protection
// ************************************************
CURR:PROT:UNLink 2
CURR:PROT:LINK?2
// queries the combined overcurrent protection of the selected channel
//response 0
[SOURce:]CURRent:PROTection:CLEar
[SOURce:]CURRent:PROTection:DELay:INITial
[SOURce:]CURRent:PROTection:DELay
[SOURce:]CURRent:PROTection:LINK
[SOURce:]CURRent:PROTection:TRIPped?
[SOURce:]CURRent:PROTection:UNLink
[SOURce:]CURRent:PROTection[:STATe]
[SOURce:]CURRent:PROTection:CLEar
Resets the OCP state of the selected channel. If an OCP event has occurred before,
the reset also erases the message on the display.
Example:
User Manual 1178.8736.02 ─ 06
........................................................................... 119
................................................................ 120
.......................................................................... 120
.............................................................................120
..................................................................... 121
......................................................................... 121
.........................................................................121
See
Example "Configuring fuses"