The R&S NGL/NGM displays the "Trigger" dialog.
3. Select the required trigger action.
●
●
●
4. Select "Set" to configure the trigger action.
6.2.3.4
Output Mode
1. Select the "Output Mode" menu item to set the output mode. See
current"
The R&S NGL/NGM displays the "Output mode" dialog.
2. Select the required output mode.
●
●
●
Note: If "Sink" mode is selected, the source current is limited to 1 mA. Otherwise, if
"Source" is selected, the sink current is limited to 1 mA.
3. Select "Set" to configure the trigger condition.
User Manual 1178.8736.02 ─ 06
"Output On": If triggered, the selected output channel is turned on.
"Output Off": If triggered, the selected output channel is turned off.
"Gated": If triggered, the selected output channel is gated.
on page 24.
Auto: Depending on the voltage across the output terminal, the R&S NGL/NGM
goes into sink or source mode.
Sink : The R&S NGL/NGM goes to sink mode, current flows into the instru-
ment. On display, the current shown negative current.
Source: The R&S NGL/NGM goes to source mode, current flows out from the
instrument.
Instrument Functions
Activating the Channels Output
"Source and sink