R&S NGM202 User Manual page 24

Hide thumbs Also See for NGM202:
Table of Contents

Advertisement

®
R&S
NGL200/NGM200
13 = Ethernet (LAN) connector
14 = Optional IEEE-488 (GPIB) interface
15 = Digital I/O connector
AC inlet with fuse holder and voltage selector (8)
Main supply cord
Do not use detachable mains supply cord with inadequate rating.
The power cable must be plugged in before signal circuits can be connected. Do not
use the product if the power cable is damaged. See
Instrument",
The built-in voltage selector selects the mains voltage between 100 V, 115 V and 230
V. All voltage settings are using the same fuse rating.
Channel connectors (9, 10)
Output terminals
Either the output terminals at the front panel or those at the back panel can be used.
Using both terminals at the same time can cause instrument malfunction.
Digital voltmeter (DVM)
The DVM+ and DVM- pins on the channel connector are available only with R&S NGM
power supply series equipped with option R&S NGM-K104 (P/N: 3643.9927.02).
The channel connectors contain both output ("F+", "F-") and sense ("S+", "S-") connec-
tions. Connector for "Ch 2" is only available in the NGL202, NGM202 models.
USB connectors (11, 12)
The USB host connector (Type-A) can be used for mass storage devices or an external
mouse like the USB connector at the front panel.
The USB device connector is a Type-B connector for remote control operation.
Ethernet connector (13)
10/100 Ethernet port for remote control operation via the local area network.
For a detailed description on the connection setup, see
nection",
User Manual 1178.8736.02 ─ 06
on page 22 for more information.
on page 77.
Getting Started
Instrument Tour
Chapter 4.2.2, "Switching On the
Chapter 6.15.1.1, "LAN Con-
21

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ngm200 seriesNgl200 seriesNgl-com2aNgl201Ngl202Ngm201

Table of Contents