NGL200/NGM200
Example:

7.3 Display Commands

The DISPlay subsystem contains the commands for display functions, which do not
affect signal generation directly.
DISPlay:BRIGhtness
DISPlay[:WINDow]:TEXT:CLEar
DISPlay[:WINDow]:TEXT[:DATA]
DISPlay:BRIGhtness <brightness>
DISPlay:BRIGhtness?
Sets or queries the display brightness.
Parameters:
<brightness>
Example:
DISPlay[:WINDow]:TEXT:CLEar
Clears the text message box on the front display.
Usage:
DISPlay[:WINDow]:TEXT[:DATA] <string>
Displays a text message box on the front display.
Setting parameters:
<string>
Example:
Usage:

7.4 Trigger Commands

The TRIGger subsystem contains the commands for signal triggering.
User Manual 1178.8736.02 ─ 06
SYSTem:TIME 12, 30, 59
SYSTem:TIME? -> 12, 30, 59
Returns system time.
........................................................................................................ 93
........................................................................................ 93
....................................................................................... 93
Displays brightness for the instrument.
Range:
0.0 to 1.0
Increment: 0.1
*RST:
0.8
DISPlay:BRIGhtness 0.5
DISPlay:BRIGhtness? -> 0.5
Returns the display brightness value.
Setting only
Text message for display.
DISPlay:TEXT "Instrument Test"
Setting only
Remote Control Commands