R&S NGM202 User Manual page 123

Hide thumbs Also See for NGM202:
Table of Contents

Advertisement

®
R&S
NGL200/NGM200
Usage:
[SOURce:]CURRent:PROTection:DELay:INITial <duration>
Sets the initial fuse delay time once output turns on.
Parameters:
<duration>
Example:
[SOURce:]CURRent:PROTection:DELay <duration>
Sets the fuse delay time.
Parameters:
<duration>
Example:
[SOURce:]CURRent:PROTection:LINK <arg0>
[SOURce:]CURRent:PROTection:LINK? <arg0>
Sets or queries the fuses of several selected channels (fuse linking).
Parameters for setting and query:
<arg0>
User Manual 1178.8736.02 ─ 06
Event
numeric | MIN | MAX
numeric
Numeric value for initial fuse delay.
MIN
Min value for initial fuse delay.
MAX
Max value for initial fuse delay.
Range:
0.00 to 60.00
*RST:
0
Default unit: s
See
Example "Configuring fuses"
numeric | MIN | MAX
numeric
Numeric value for the initial fuse delay.
MIN
Min value for initial fuse delay.
MAX
Max value for initial fuse delay.
Range:
0.00 to 10.00
*RST:
0
Default unit: s
See
Example "Configuring fuses"
1 | 2
Remote Control Commands
Configuration Commands
on page 118.
on page 118.
120

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ngm200 seriesNgl200 seriesNgl-com2aNgl201Ngl202Ngm201

Table of Contents