Appendix 2 Error Messages and Remedial Actions
Message
CSV auto save has been set.Un-
239
able to set the specified record in-
terval.
The timer stop time has been
240
passed.
10ms recording interval cannot be
241
set due to too many channels.
20ms recording interval cannot be
242
set due to too many channels.
Recording interval has been
243
changed.
Auto-save has been changed from
244
CSV to waveform.
301
No waveform data.
302
Confirm the A-B cursor position.
303
No numerical calculation data.
304
Invalid event mark.
401
File processing error.
402
Cannot load this file.
403
Insert a Media.
404
Directory full.
405
Not enough capacity.
Remedial Action
Select non-CSV auto saving to measure with the
specified recording interval.
Change the timer stop time or cancel the timer.
Measurement with 10 ms recording interval is avail-
able when using UNIT1-1 to -15. Disable (set to Off)
channels on other input units (2 to 4).
Measurement with 20 ms recording interval is avail-
able when using UNIT1-1 to UNIT2-15. Disable (set
to Off) channels on other input units (3 and 4).
The recording interval has changed due to expand-
ed channel usage.
The saving method was changed to Waveform (bi-
nary) because the selected recording interval is too
short for CSV saving. To enable CSV saving, select
a longer recording interval or reduce the number of
channels used.
"Automatic Saving (Waveform Data and
See:
Numerical Calculation Results)" (p .127) ∗2
No waveform data is preset, so data saving and cal-
culations cannot be performed. Press
start acquiring measurement data.
The A/B cursor positions are invalid (out of wave-
form range). Check the cursor positions.
Execute numerical calculation
Try moving to an event mark that is not in memory.
An unexpected fault occurred while processing a file
on the removal storage. Turn the HiLogger off and
back on.
The format of the file is incompatible with the HiLog-
ger, or the file is corrupt.
Refer to "6.1 About Saving and Loading Data" (p.
123) for loadable file formats.
A CF card or USB flash drive is not present. Insert
storage media.
No more files can be created because the directory
is full. Check or perform the following:
1. System reset (p. 153)
2. Verify that the removal storage is specified by
Hioki (p. 47)
3. Reformat the removal storage (p. 49)
If the message persists after performing the above,
either the removal storage or HiLogger may be dam-
aged. Contact your dealer or Hioki representative.
The removal storage as insufficient space to save
the file. Delete files to increase free space, or re-
place the removal storage.
START
to
(p. 155)