Appendix A
2511-
Print job (name) has been cancelled due to insufficient disk space. Delete unnec-
essary files and restart the report.
Cause:
Action:
2512-
Print job (name) has been cancelled.
Cause:
Action:
2513-
Unable to read the calibration file (file name).
Cause:
Action:
2514-
Unable to write the calibration file (file name).
Cause:
Action A:
Action B:
2515-
Warning: Changing the calibration information will affect the performance of the
instrument. Only qualified service personnel should do this. Do you wish to
proceed?
Cause:
Action:
A-12
The disk drive does not have enough space for the temporary file
required by the Windows Print Manager. Therefore, printing of the
requested report has been canceled.
Delete unnecessary files from the disk. You will require at least five
megabytes of free space for normal operation.
The requested print job was canceled at your request.
None required.
You selected an invalid calibration file or one that cannot be read.
Be sure the media containing the calibration file has no problems.
An attempt to Save calibration data has failed due to possible media
problems.
Be sure the media you want to Save the file to has no problems.
Choose an alternate media to Save the calibration data.
You have started the process of performing a calibration operation.
Calibration operations should only be done by, or under the direction
of, qualified service personnel.
DVVA II 4000
400-42803-01 (Rev B) - July 2017
Need help?
Do you have a question about the DVVA II 4000 and is the answer not in the manual?
Questions and answers