Micromeritics DVVA II 4000 Operator's Manual page 225

Dynamic void volume analyzer
Table of Contents

Advertisement

DVVA II 4000
2506-
(sample file) Output device (name) is not installed. Printing cannot be
accomplished.
Cause:
Action:
2508-
(sample file) Overlay file (name) was not found. It will not be included in the
reports.
Cause:
Action:
2509-
(sample file) Error opening file (name): (error). Reports cannot be produced.
Cause:
Action:
2510-
(sample file) Error parsing reports from file (name). Reports cannot be produced.
Cause A:
Action A:
Cause B:
Action B:
Cause C:
Action C:
400-42803-01 (Rev B) - July 2017
The selected output device is not installed in Windows.
Install the device using the Control Panel > Printers operation.
The specified overlay file could not be found.
Ensure that the file specified as an overlay does exist.
An error occurred while the program was opening a file necessary to
the report operation.
Use the name given in the error message to investigate. Contact your
service representative if you continue to receive this error message.
One or more data entry fields in the sample file may contain an invalid
character (such as a single quote or double quotes).
Review the data entry fields (for example, the Sample field) and
remove the invalid character.
The system was unable to create the usual temporary files during the
report, possibly due to insufficient disk space.
Check the space available on the hard disk.
An internal processing error occurred.
Contact your service representative.