Appendix A
2449-
This field does not contain a valid file specification.
Cause:
Action:
2450-
Sample Defaults may not be edited while this operation is in progress. Do you wish
to save and close the Sample Defaults edit session?
Cause:
Action:
2452-
Attempt to write MICATTR.DIR in read only mode. (file name)
2453-
Attempt to append MICATTR.DIR in read only mode. (file name)
Cause:
Action:
2454-
Too many selections for a print-to-file operation. Only the first (number) selections
will be processed. Please reselect the remainder.
2455-
Too many selections for an export-to-file operation. Only the first (number) selec-
tions will be processed. Please reselect the remainder.
Cause:
Action:
A-4
You entered an invalid file name.
See the description of file naming conventions in your Windows man-
ual and re-enter the name.
You are in the process of initiating an automatic analysis (an analysis in
which sample files are created using the defaults) while editing the
defaults.
Finish your edit session of the defaults and close the dialog. Then
restart the automatic analysis.
The Read-Only attribute is turned on in the application's
MICATTR.DIR file (this file exists in each folder containing sample or
parameter files).
Use Windows Explorer to access the folder containing the
MICATTR.DIR file and disable the Read-Only option.
You selected too many files for this operation.
Select only the number of files specified in the message.