Troubleshooting

Full Power-On Procedure

The power-on procedure is to be performed after the instrument has been shut down for an
extended period of time or after a loss of power.
1. Make sure the instrument and computer are in the following state before continuing:
a. Computer is turned OFF.
b. The instrument MAIN POWER switch (on back panel) is in the OFF position.
c. The instrument Power switch on the front panel is in the OFF position.
d. The USB cable is disconnected from the instrument.
2. Place the instrument MAIN POWER switch on the back panel to the ON position.
3. Connect the USB cable to the instrument.
4. Turn on the computer.
5. If required, allow Windows to install the drivers for any new hardware found. Refer to
"Power-On Troubleshooting" on page 9-6.
6. Place the instrument front panel Power switch in the ON position.
7. Start the DVVA II 4000 software.
9-4
DVVA II 4000
400-42803-01 (Rev B) - July 2017