DVVA II 4000
Use Prior Blank
Measure Residual Volume
Time Between Cycles
400-42803-01 (Rev B) - July 2017
Skips the blank analysis automatically performed before each
analysis and uses the data from the previous blank analysis.
When using this option for your sample analysis, the height of
sample material in the sample chamber must be higher than that
used in the blank analysis.
Allows you to measure the void volume of the compressed
sample. At the end of the compression cycle, the upper piston
withdraws and a single-point measurement is collected. This
point defines the top of compressed sample and is used to
calculate the residual void volume.
Enter how many minutes to wait after the completion of the
compression cycle before taking the single-point measurement.