7. Enter values in the Skeletal Density and Bulk Density fields that are typical of your sample.
These values, like all other values, can be edited in the sample file.
8. Select the appropriate material from the Material Properties drop-down list. The Material
type is used to determine which carbon property models are used for this particular carbon.
For example, a different equation is used to predict COAN for Tread/Hard carbons than for
Carcass/Soft carbons. If you select Other, the COAN is not calculated.
9. If you plan to report statistical process control (SPC) information, enter appropriate
information in the User Parameter fields. These user-definable parameters enable you to
track statistical process control data; refer to Sample Defaults, page
information.
10. Click the Analysis Conditions tab. Choose the analysis conditions appropriate for your most
commonly analyzed material.
11. Click the Report Options tab. Choose the desired reports, using
click
Save
12. Click
Close
400-42803-01 (Rev B) - July 2017
.
to close the dialog.