Resetting A Calibration; Verifying A Calibration - Emerson Rosemount Xstream Series Instruction Manual

Gas analyzer
Hide thumbs Also See for Rosemount Xstream Series:
Table of Contents

Advertisement

X-STREAM
7-3-5

Resetting a Calibration

CANCEL calibration!
START calibration!
ZeroGas
CO2.1
RESET calibration..
CO2.1
RESET calibration
Are you sure?
No!
Yes!
7-3-6

Verifying a Calibration

Control..
Zero calibration..
Span calibration..
Apply gas..
7-30
7-3
Calibration Procedures
0.000 ppm
0.200 ppm
0.200 ppm
In case a wrong configuration was detected
after calibration was performed (e.g. wrong gas
connected) there is an option to restore the last
user saved calibration data:
Within the screen where to start the calibration
(either for span or zero) enter line 4:
A new screen appears with the option RESET
CALIBRATION.. Pressing the ENTER key in
this line results in a prompt for confirmation.
Choosing YES replaces the current calibration
data with the last calibration data (from
UserData;
5-4-3-8 "SAVE-LOAD", page
5-45).
For instruments without internal and/or
external valves simply apply either span or zero
calibration gas to the sample gas inlet. If the
calibration still is proper, the reading on the
measurement screen should show the related
value.
For instruments with internal and/or external
valves follow the procedure below:
Starting from the measurement screen press
the DOWN key to open the MAIN MENU, enter
the CONTROL.. menu.
Enter the last line (APPLY GAS..)
Dual channel instrument:
Optional gas component selection menu -
Select the component to be setup
Emerson Process Management GmbH & Co. OHG
Instruction Manual
HASAxE-IM-HS
05/2006