Display/Ao Setting; Setting Of Instantaneous/Average Value - Fuji Electric FE ZSS Instruction Manual

Cross stack laser gas analyzer
Table of Contents

Advertisement

Make a setting of "Inst./Avg." or "Wet/Dry" which displays the measurement value or its external output
value displayed on the "Measurement" screen. In case of 2 range specification, range setting is performed.
Explanation of setting items
"Inst./Avg."
"Wet/Dry"
"Range Control" : In case of 2 range specification, the following setting can be selected.
Select "Display/AOSetting" from the "Menu" screen, and press the
Menu
Zero Calibration
Span Calibration
Alarm Log
Alarm Setting
Output Hold

6.6.1 Setting of instantaneous/average value

Make a setting of "Instantaneous value" or "Average value" for each measurable component. The setting is
reflected on the display on the "Measurement" screen or external output such as analog output.
O
conversion value can be arbitrarily selected even if the previous measurement value is set as instantane-
2
ous value (or average value).
When the average value is set, moving average processing is performed at the time set at "Average Period"
on the "Parameter Setting" screen.

6.6 Display/AO setting

: Make a setting of "instantaneous value" or "average value" for each measur-
able component.
: Make a setting of "Wet" or "Dry" for each measurable component. When
"Dry" is set, H
O setting of the "Analog Input" is required.
2
"Auto"
"Range1", "Range2" : Fixed range
"Remote"
2006-04-01
12:00:00
Parameter Setting
Display/Output
Input Signal
AO Select
Password
*****
: Automatic range switching
: Analog input switching
NT
Display/AOSetting
Inst.
Gas
Inst.
HCℓ
cv
Inst.
HC ℓ
NT
- 54 -
key.
Wet
Range Control
Avg.
Dry
Range1 0 ~ 50.0 ppm
Wet
Remote
Range2 0 ~ 1000ppm
Range1 0 ~ 50.0 ppm
Dry
Range2 0 ~ 1000ppm
2007- 4 -1
12:00:00
INZ-TN1ZSS-E