Battery Power Level Display On Handpiece; Operation; Selection Of Exposure Time - 3M Elipar DeepCure-L Manual

Led curing light
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
SEITE 6 - 148 x 210 mm - 44000118149/02 - SCHWARZ - 20-113 (kn)
The green status light on the handpiece blinks while
the battery is charging. The green status light glows
steadily when the device is fully charged. As a safety
precaution, light-curing is not possible while the
device is charging.

Battery Power Level Display on Handpiece

Status LED
Handpiece
without charger connected
Steady green
Handpiece
light
ready for

operation

Flashes green
––
Steady red light Low battery
charge
Flashes red
Battery fully
discharged,
exposure
cycle will be
completed or,
if in continuous
mode, stopped
6
Status LED
Operating status
Charger
Charging has
been completed
Battery is being
charged
Problem during
charging
Problem in
charging, battery
is defective
or cannot be
charged
Operation

Selection of Exposure Time

{
LED showing
exposure times
Exposure time options: 5, 10, 15, 20 sec, continuous
mode (120 sec), tack-cure mode (1 sec).
E
See the instructions for use for the specific dental
material when selecting the exposure time.
E
The indicated exposure times assume that the light
guide is held at the exact position over the material
being polymerized.
E
If the distance between the light guide and the
restoration is increased, the exposure time must be
adjusted accordingly because the light intensity
weakens (see graph).
100
90
80
70
60
50
0
1
2
3
4
Distance (mm)
Select the exposure time by pressing the TIME button.
- The selected exposure time is indicated by the
4 green LEDs.
- Each time the button is briefly pressed, the
setting advances to the next (higher) value. All
4 green LEDs will be turned on for a setting of
20 sec. Pressing the button again will turn off all
of the LEDs and enable the continuous mode.
- The display advances through the available
settings, if the button is kept depressed.
- While exposure is ongoing, the button for selec -
tion of the exposure time is inactive.
START button
TIME button
5
6
7
8
9 10

Advertisement

Table of Contents
loading