Troubleshooting

Error
Cause
E
The status display on
The remaining battery
the handpiece glows
charge is adequate only for
red steadily.
typical 5 10-sec exposure
cycles.
E
The status display of
The battery lacks sufficient
the handpiece flashes
charge.
red.
E
The ongoing exposure
is interrupted (light off
signal is emitted)
followed by a 2 sec-
error signal; the hand-
piece switches to
"sleep" mode and
resists further activa-
tion.
The status light on the
Charging problem. The
handpiece blinks red
battery is defective or at the
whenever the hand-
end of its useful life.
piece is connected to
E
the charger.
The handpiece has
There is not enough charge
not been used for a
in the battery to turn on the
long time and now it
handpiece.
cannot be turned on.
E
The handpiece does
Software crash possible.
not respond to the
E
pressing of either
button.
While the handpiece
The handpiece is connected
is connected to the
to the charger. Light-curing
charger, pressing
is prevented as a safety
the START button
precaution.
will not start the light
E
emission.
Solution
Connect the handpiece
to the charger and
recharge the battery.
Connect the handpiece
to the charger and
recharge the battery.
Contact 3M
Service.
Connect the handpiece
to the charger and
recharge the battery.
Plug the included
charger into an outlet
and connect it to the
handpiece. This causes
the light-curing device
to reset itself.
Disconnect the charger
from the handpiece and
restart the light emission.
Error
Cause
E
The light emission
The information about inter-
does not start when
mittent operation under the
the START button
header "Technical data -
is pressed; an error
handpiece" has not been
signal sounds for
observed.
2 sec.
The handpiece has become
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
E
Allow the handpiece to
cool for 3 minutes, then
start the next exposure
by pressing the START
button.
During light emission
The information about inter-
in continuous mode,
mittent operation under the
an error signal sounds
header "Technical data -
for 2 sec, the emission
handpiece" has not been
is stopped, and the
observed.
handpiece changes
The handpiece has become
to sleep mode.
overheated in the course
of use. The handpiece can
be used again once it has
cooled down.
E
Allow the handpiece to
cool for 3 minutes, then
start the next exposure
by pressing the START
button.
E
The light intensity is
Clean the light guide and
too low
the protecting glass in the
light guide mounting hole
(please refer to "Cleaning
the Light Guide").
E
The dental material
Clean the light guide and
does not cure
the protecting glass in the
completely.
light guide mounting hole
(please refer to "Cleaning
the Light Guide").
E
Check to see that the
correct light guide has
been attached.
E
The light guide cannot
The light guide is not
be attached to the
designed for use with the
handpiece.
Elipar DeepCure-L.
Solution
9