Cleaning And Disinfecting Charger, Handpiece And Glare Shield; Storage Of The Handpiece During Extended Periods Of Non-Use; Return Of Old Electric And Electronic Equipment For Disposal; Customer Information - 3M Elipar DeepCure-L Manual

Led curing light
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
SEITE 12 - 148 x 210 mm - 44000118149/02 - SCHWARZ - 20-113 (kn)
• Maximum sterilization temperature 135 °C (275 °F)
• The following cycles were validated for the light
guide:
Fractional
prevacuum
Tempera-
132 °C
134 °C
ture
(270 °F)
(273 °F)
Sterilization
4 min
3 min
time
Drying time 30 min
30 min
Check
Before using the light guide again, check it for dam-
aged surfaces, discoloration, and contamination; do
not use damaged light guides. If the light guide is still
contaminated, repeat all reprocessing steps.
Cleaning and Disinfecting Charger, Handpiece and
Glare Shield
• Cleaning and disinfecting solutions must not enter
the unit.
• Wipe the surfaces with ready-to-use disinfectant
wipes* (e.g., CaviWipes™ based on the active sub-
stances alcohols and quaternary compounds (QAV)
or equivalent disinfection agents). Subsequently in-
spect the surfaces for visible impurities. If any impu-
rities are visible, repeat the cleaning procedure.
• Do not use solvents or scouring agents, as they
may damage the device components.
• Wipe the device surfaces with a tuberculocidal dis-
infectant wipe* for disinfection and keep it wet ac-
cording to the disinfection agent manufacturer's di-
rections for use, particularly observing the contact
time.
• Disinfectants must not be sprayed directly onto the
device surface.
• Wipe the device surfaces with a soft cloth moist-
ened with water (drinking water quality as a mini-
mum) and dry with a soft, lint-free cloth. If you do
not rinse well with water, the components will be
damaged.
• If necessary, ask the manufacturer of the disinfec-
tant if its constant use will damage plastic surfaces.
Clean the glare shield with a soft and fluff-free cloth.
Beware of scratches!
Ensure that the charge contact pins stay dry and are
not touched by any metallic or greasy parts. Dry up
necessarily damp contact pins. Do not bend the
charge contact pins during drying. Damp charge con-
12
tact pins cause an operating error (malfunction mes-
sage: the LED showing the operating status of the
charger glows red steadily).
*CaviWipes™, Metrex, were used for the validation of
Gravity
the cleaning and disinfection procedure.
displacement
Storage of the Handpiece during Extended Periods
121 °C
of Non-Use
(250 °F)
E
If the handpiece will not be used for a number of
weeks - e.g., during vacation - charge the battery
20 min
beforehand or connect the handpiece to the charger
for this time. A safety switch within the battery
prevents a total discharge.
30 min
Discharged or nearly discharged batteries must be
recharged as soon as possible.
Return of Old Electric and Electronic
Equipment for Disposal
1. Collection
Users of electric and electronic equipment are required
to collect their old equipment separately from other
waste in accordance with the regulations of the specific
country. Old electric and electronic equipment must not
be disposed of with unsorted household waste. This
separate collection is a prerequisite for recycling and
reprocessing as an important method for preserving
environmental resources.
2. Return and Collection Systems
When your Elipar DeepCure-L is no longer usable,
do not dispose of the device with household waste.
3M Deutschland GmbH has set up special disposal
facilities to handle the equipment. Details about the
procedure for the specific country can be obtained
from the pertinent 3M subsidiary.
3. Removing the Battery for Disposal
To dispose of the battery, disconnect the Elipar
DeepCure-L LED curing light from the charger, remove
the two screws on each side of the charging socket
and push the lower half of the housing towards the
back, away from the upper half of the housing. Use a
suitable tool to cut the connecting wires between the
battery and the circuit board and remove the battery
for disposal in accordance with 1. and 2.

Customer Information

No person is authorized to provide any information
that deviates from the information provided in this
instruction sheet.