Cryogenic Systems Operation
Inspect the interior of the cryogenic arm of the probe
for foreign objects and extra bushings. Remove any
foreign materials and bushings.
Verify that the brass stinger tip is screwed on tightly
and is not loose.
4
Install the plastic bushing on the stinger with the flat
face towards the CryoBay (curved face towards the
stinger tip).
Install the rippled plastic sleeve over the stinger, as seen
5
in
Figure
4. The sleeve may remain inside the probe after
first use.
Figure 4
CCC Stinger with plastic bushing and plastic sleeve
Position the flexible cryogenic transfer line, as seen in
6
Figure
5, of the refrigerator assembly in line with the
probe side arm in preparation for insertion of the stinger
into the probe side arm.
Cryogenic Operation
Plastic Sleeve
2
Plastic Bushing
19