Agilent Technologies Cryogenic User Manual page 114

Table of Contents

Advertisement

8
Appendix
114
Disable the sample catcher release lever and release the
13
turbine sample. DO NOT disable the sample catcher or
press the release button if the eject air is not on.
Insert the turbine and sample into the probe from VnmrJ
14
or VNMR using the insert button or insert command.
15
Enable the sample catcher.
16
Open the CryoBay sample eject pneumatic valve. The
sample should float to the top of the upper barrel.
Verify that the lever of the sample catcher has engaged
17
the turbine and sample.
18
If necessary, adjust the needle valve on the eject line
until the sample lifts to the top of the upper barrel and
the sample catcher engages the sample.
Close the CryoBay sample eject pneumatic valve.
19
20
Repeat step 12 through step 19 as necessary to properly
adjust the eject air.
Cryogenic Systems Operation