Host Computer Failure - Agilent Technologies Cryogenic User Manual

Table of Contents

Advertisement

7
Unexpected Events

Host Computer Failure

102
The CryoBay Monitor program on the host computer is the
user interface to the CryoBay internal computer board. The
CryoBay internal computer board operates the CryoBay and
compressor.
If the host computer fails, the CryoBay will continue with
the last sequence of operations that it was commanded to
do, such as, Start. It will continue until it is in the
Operating state. It will continue in the Operating state until
an error occurs or a new command is received from the host
computer.
When the host computer is operating again, open the
CryoBay Monitor and click the Connect button. See
"CryoBay Monitor"
on page 65.