4
Principles of Operation
60
Table 3
Start - Initial state: Idle
State
CryoBay order of operations/conditions
Pump out CryoBay vacuum system until low vacuum.
Pump out CCC vacuum chamber until low vacuum.
Open probe valve (popup button will appear on probe valve).
Cooling
Pump out probe until low vacuum.
Purge helium and fill helium for cold loop (repeated 5 times,
fill to 215 psi (1.55 mPa)).
Turn on compressor.
Cool probe until probe reaches set-point temperature.
Operating
Temperature controller regulates probe internal
temperature.
Ion pump on.
Turbo pump off.
Cryogenic Systems Operation