Unexpected Events

CCC has reached room temperature

104
Click Send CryoBay.
4
The system will resume the Warming state.
5
Enter <SETWARM>
on command line
Figure 41 <SETWARM> in command line
1
If there has been a single- phase power failure or a host
computer failure, re- open the CryoBay Monitor program
and click the Connect button. See
Monitor (Vertical Panel)"
When the system is power cycled (returning to nominally
2
Idle state), it is normal for the CryoBay sample eject air
to come on. Turn off the sample eject air by pressing the
Stop button in the Cryobay Controls window.
If the CryoBay sample eject air was manually turned off
3
by closing the CryoBay sample eject air needle valve, it
will be necessary to adjust the CryoBay sample eject air.
See
"CryoBay Sample Eject Air and Sample Catcher" on
page
110.
The system may be restarted using the standard Start
4
procedure. Do not forget to review
Cautions and Warnings"
"Opening the CryoBay
on page 66.
"Final System Checks,
on page 24.
Cryogenic Systems Operation