Retightening Swagelok Connectors - Agilent Technologies Cryogenic User Manual

Table of Contents

Advertisement

Retightening swagelok connectors

Cryogenic Systems Operation
Do not disassemble the Swagelok.
Insert the tube with the ferrule attached into the fitting
1
body, as seen in
2
Seat the tube and ferrule into the fitting body.
Finger- tighten the nut until it is in the original position.
3
The resistance will increase when the nut is at its
original position.
4
Tighten the nut slightly from the position of resistance.
Insert tubing with the ferrule
until it seats
Body nut
Finger-tighten the Swagelok
until resistance increases
then tighten slightly with a wrench
Figure 48 Retightening swagelock connectors