8
Appendix

Swagelok and Aeroquip Fittings

Swagelok connector installation

116
Do not disassemble the Swagelok fitting.
Trim both ends of the copper tubing using a tube cutter.
1
Do not use a file, saw, or clipper. Both ends of the
copper tubing must be round and not crimped.
Insert the tubing into the Swagelok, see
2
seat the tubing against the shoulder of the fitting.
Tighten the nut by hand until the tube will not rotate
3
freely in the fitting. If the tube cannot be rotated, tighten
the nut finger- tight, and then use a wrench to tighten the
nut an additional 1/8 of a turn.
Scribe a line on the nut at the 6 o'clock position.
4
Hold the body nut with a wrench and, with another
5
wrench, tighten the nut 1 1/4 turns.
Insert tube into Swagelok and
tighten the nut finger-tight
Body nut
Figure 47 Swagelok installation
Scribe line at 6 o'clock position
Finger-tighten then tighten the
Swagelok nut 1 1/4 turns
Cryogenic Systems Operation