5
Troubleshooting and Error Handling
Power Failure on AI
Possible Causes
The Well-plate Sampler was switched off
The 24V power was disconnected from the
Well-plate Sampler or from the Automation
Interface
The 24V power cord is damaged
The internal 24 V power circuit has failed
Well-plate Sampler Thermostat Enabled Error
Possible Causes
The Well-plate Sampler thermostat was switched
on
84
Actions
Switch on the Well-plate Sampler
Reconnect the 24V power cord
Replace the 24V power cord
Verify functionality using Automation Interface
test in diagnosis
Actions
The Well-plate Sampler thermostat should not be
used while the Automation Interface is installed
1100 Series Sample Capacity Extension