7. Enter the concentration of the low calibration standard for Cal B. This
can be the same standard used for calibration of the High Cal A or can
use nitrogen as a zero standard.
C a l
i b r a
L o w
C a
∨
Figure 5-36.
8. This step will latch the detector response and associate it to the High
Cal A value.
C a l
i b r a
R e c a
l
i b r a t e
∨
Figure 5-37.
9. This step will associate the response of the detector to the Low Cal A
value.
C a l
i b r a
R e c a
l
i b r a t e
∨
Figure 5-38.
10. This step moves the calibration line so the reading matches the user-