Table A–25. Oven settings: Operations Page > Monitor 1
Description
The control mode currently in effect. Possible displays listed below.
The current heat output level (%).
The current cool output level (%).
The set point currently in effect (°C).
The current filtered process value using the control input (°C).
Table A–26. Oven settings: Operations Page > Control Loop 1
Default
Display
Off
Auto
Manual
Software
Auto
90%
No
110.0°C
25.0°C
14.0°C
2.0°C
14.0°C
2.0°C
180
0
0.0°C
0.0%
SOLA II Flare User Manual
Appendix A
Watlow EZ Zone PM6 Series
Comment
The loop will use automatic control.
The percentange of the current set
point that will be used by autotune.
The set point the controller will
automatically control to.
The closed loop set point that can be
triggered by an event state.
The PID proportional band for the heat
outputs.
The process value needs to be this far
into the On region before the output
turns on.
The PID proportional band for the cool
outputs.
The process value needs to be this far
into the On region before the output
turns on.
The PID integral (seconds).
The PID derivative (seconds).
The offset to the proportional band.
The fixed level of output power when
in manual (open loop) mode.
A-15