Alarms Setup

Thermo Scientific
Review the following for instructions pertaining to alarm setup.
1. Access the menu items by pressing Enter.
C o n f i g u r a t i o n
A l a r m s
Figure 5-17.
2. To set up a Chamber Flow alarm, enter the set point at which the low
flow alarm for the flow measured at the detector chamber in the PUVF
should activate. The value can be from 0 to 1000 cc/m.
A l a r m s
C h a m b e r
Figure 5-18.
3. To set up a Lamp Voltage alarm, enter the high point lamp voltage at
which the alarm should activate. The value can be from 0 to 1200 V.
A l a r m s
L a m p
V o l t a g e
Figure 5-19.
S e t u p
E n t
S e t u p
F l o w
A l a r m
E n t
S e t u p
A l a r m
E n t
SOLA II Flare User Manual
Chapter 5
Configuration
E x i t
E x i t
E x i t
5-6