Com Ports RX TX

Thermo Scientific
Following are the menu items within the View Parameters menu.
Table 7-3.
Menu Item
Description
TS Reading
The concentration reading.
Cal A Response Factors
The slope (kHz/ppm or kHz/ppb sensitivity) and the offset (calculated at 0
concentration).
The slope (kHz/ppm or kHz/ppb sensitivity) and the offset (calculated at 0
Cal B Response Factors
concentration).
Working Average Sec
When switching from the normal average to the fast average and back to
normal, the average time used can be observed from this menu item.
AVG PMT signal kHz
The average of the raw detector signal.
The frequency proportional to the PMT output normalized to the gain of
Normalized PMT kHz
100; also displays the gain currently selected.
PMT Signal kHz
The actual frequency proportional to the PMT output; also displays the
gain currently selected.
PMT Voltage
The input voltage to the PMT.
Internal Temperature
The temperature inside the electronics enclosure.
Chamber Temperature
The temperature inside the reaction chamber.
Chamber Flow
The flow through the reaction chamber.
Chamber Pressure
The pressure sensed at the reaction chamber.
Lamp I Alm Counter
The frequency proportional to the lamp intensity.
Lamp Voltage
The lamp input voltage.
This menu displays activity for Coms 1–4.