The Recalibrate High
Function

Adjusting the Reading

Thermo Scientific
8. Press the Yes button.
9. Manipulate the valve on the sample conditioning system to allow the
sample to once again flow into the instrument.
Use this function if manually introducing the high calibration standard to the
analyzer.
1. Go to Configuration > Calibration Setup > High Cal A Value. Set the
high calibration value to match the concentration of the standard used.
Exit the menu.
2. Connect the calibration standard to the unit, and let it flow through the
injection valve.
3. Let the reported concentration reach a new value and stabilize.
4. Go to Configuration > Calibration Setup > Recalibrate A High. Press
Yes to accept the new high calibration.
At this point, the current average of the detector signal is latched to correspond
to the high calibration value concentration.
If the reported value of the SOLA II Flare does not match a known sample
value or the standard flowing through the instrument, you can adjust the
reading. To do so, go to Configuration > Calibration Setup > Adjust Reading.
The current reading will be displayed in an editing screen so that you can
modify the reading to match the desired reading.
Note Do not use this function if the difference between the expected reading
and the analyzer's reported reading is significant. The Adjust Reading function
should only be used for making minor corrections to the calibration. Using this
function improperly can hide malfunctions in the unit that need to be
corrected. ▲
Configuration
SOLA II Flare User Manual
Chapter 5
5-12