4. To set up a Lamp Intensity alarm, enter the low-point lamp frequency
at which the alarm should activate. The value can be from 0 to 50,000
Hz.
A l a r m s
L a m p
I n t e n s i t y
Figure 5-20.
5. To set up a Lamp Voltage Rate of Change alarm, enter the rate of
change in lamp voltage at which the alarm should activate. The value
can be from 1 to 50 V per 30 seconds.
A l a r m s
L p
V
R a t e
Figure 5-21.
6. Set up a Chamber Temperature alarm by entering the +/- deviation
from 45°C at which the alarm should activate.
A l a r m s
C h a m b e r
Figure 5-22.
S e t u p
E n t
S e t u p
o f
C h a n g e
E n t
S e t u p
T e m p .
E n t
SOLA II Flare User Manual
Chapter 5
Configuration
A l a r m
E x i t
E x i t
A l a r m
E x i t
5-7