Parameter
Display
Software
Heat Algorithm
Cool Algorithm
TRU-TUNE+ Enable
TRU-TUNE+ Band
TRU-TUNE+ Gain
Autotune Aggressiveness
Peltier Delay
User Failure Action
Input Error Failure
Fixed Power
Open Loop Detect Enable
Open Loop Detect Time
Open Loop Detect Deviation
Ramp Action
Ramp Scale
Ramp Rate
Low Set Point
High Set Point
Set Point Open Limit Low
Set Point Open Limit High
Thermo Scientific
Table A–18. Oven settings: Setup Page > Control Loop 1
Default
Display
Software
PID
Off
No
0°C
3
Critical
0.0
User
Off
0.0%
No
240
6.0°C
Off
Minutes
1.0°C
0°C
130.0°C
-100.0%
100.0%
SOLA II Flare User Manual
Appendix A
Watlow EZ Zone PM6 Series
Comment
The method of heat control.
Critical damped aggressiveness of
autotuning calculations.
When user switches to manual mode,
controller sets output power to last
open-loop set point entered by user.
When an input error switches control
to manual mode, controller sets
output to 0%.
The minimum value of the closed loop
set point range.
The maximum value of the closed loop
set point range.
The minimum value of the open loop
set point range.
The maximum value of the open loop
set point range.
A-11