Chapter 1
Product Overview
These two features provide complete flexibility within a single analyzer system.
The lower detectable limit for the low measuring range is based upon the high
range selected; for example, if the high range is set to 1,000,000 ppm (100%)
the low range is automatically fixed at 10,000 ppm (1% of full scale). This
would dictate a lower detectable limit on the low range of 100 – 200 ppm
depending upon the final set-up used.
Figure 1-1. Typical SOLA II Flare analytical oven flow diagram
Set-up and programming of the SOLA II Flare is easy using the front panel
display. When compared to a zinc-continuous lamp, the Xenon flash lamp in
the SOLA II Flare has a cooler operating temperature. This allows for a longer
lifespan of the lamp, as well as increased sensitivity to the high energy generated
when the lamp pulses. The PUVF detector uses a 120-second rolling average
that updates the output every second, providing an excellent response time to
process changes.
Thermo Scientific
SOLA II Flare User Manual
1-2