Table 1-2.
Analog Outputs
Two 4–20 mA dc outputs, one assigned to each range
Alarm Outputs
One (1) global dry contact triggered by one or more of the following:
Low detector flow alarm
Oven/pyrolyzer temperature fault
Injection valve fault
Purge failure
Detector temperature fault
Detector lamp voltage fault
One (1) out of service dry contact triggered by:
Analyzer in validation
Suspension of analyzer
Digital Data
Dual channel RS-485 Modbus, TCP/IP Ethernet via Modbus, TCP/IP
Communications
encapsulated Modbus & RS-485 Modbus
Local HMI
Status of all analyzer parameters (e.g. pyrolyzer furnace and analytical oven
temperatures, PMT and lamp voltages, detector flow rates, etc.) and
analytical results available on the front mounted displays, push button menu
access; hazardous area classification remains intact while operating the local
display
SOLA II Flare
Complete remote control of SOLA II Flare, automatic logging of analysis
Modbus Interface
results and analytical parameters, communication to SOLA II Flare via serial
or TCP/IP encapsulated Modbus enables remote diagnostics via modem
SOLA Web Remote
Complete remote control of the SOLA II Flare, ability to download 24 hours of
Interface
analysis results and analyzer parameters, communications to the SOLA II
Flare via local area network (TCP/IP) enables remote diagnostics via modem.
The Flare SOLA Web is limited to the replica of the front display on the SOLA
II Flare.
Table 1-3.
Ambient Temperature
Input Voltage
Power
Dimensions
Zone 1 Div. 1 Configurations
Zone 2 Div. 2 Configurations
Available Certifications
54 to 104 def. F (12-40 deg. C)
Ingress Protection IP-51/dustproof and dripping water
110VAC, optional 220VAC to 110VAC Step Down Transformer
20 amp circuit recommended; 17 amp maximum during warm-up
cycle; 7-8 amps once operation temperature achieved
1581.15 mm (62.25 in) high x 647.70 mm (25.50 in) wide x 476.25
mm (18.75 in) deep
1104.39 mm (43.48 in) high x 647.70 mm (25.50 in) wide x 476.25
mm (18.75 in) deep
CSA/US: Class I, Div 2, Groups B, C & D, T3
CSA/US: Class I, Div 1, Groups B, C & D, T3
ATEX: Zone 2, 3 G Ex pz IIC T3