Thermo Scientific SOLA II Flare User Manual page 95

Sulfur online analyzer
Table of Contents

Advertisement

Thermo Scientific
22. With an air gun placed at the end of one fitting, blow out any dust or
debris that may have dropped inside the pyrolyzer quartz or glass tube.
Figure 8-14. Blow out the debris
23. Secure the pyrolyzer housing to the oven enclosure using the screws
removed earlier (two mounting screws on top of housing, two on
bottom).
24. If necessary, reconnect the wires from the oven temperature switches.
25. Reconnect the heater wires to the pyrolyzer.
26. Reconnect the thermocouple wires.
27. Reconnect the sample tubing.
28. Follow the
Initial Startup
Maintenance & Troubleshooting
procedure to restart the analyzer.
SOLA II Flare User Manual
Chapter 8
8-15