Sulfur online analyzer
Table of Contents

Advertisement

Thermo Scientific
4. Close the electronic enclosure door. Ensure that it is tightened securely.
5. Apply power to the X-Purge unit.
6. After the predetermined purge time, the X-Purge unit applies power to
the analyzer system. Typically, this takes about 10 minutes, but it may
vary depending upon the system. Refer to the system application notes
or logbook for the correct purge time.
Caution Do not open the X-Purge housing unless power is removed from the X-
Purge or the area is known to be non-hazardous. ▲
If the unit does not apply power to the analyzer system after the
required length of time, check for the following possible problems:
a. Purge pressure at the analyzer purge pressure gauge is too low.
b. Oven heater air pressure regulators set too low.
c. Oven door open or leaking air.
d. Electronic enclosure door open or leaking air.
7. Complete the remaining steps listed in the start-up procedure included
in the start-up instructions in the instrument user's guide.
Appendix F
The X-Purge System
SOLA II Flare User Manual
F-6