Maintenance By The User; Exchange Of The Measuring Tube And The Jacket Tube - Thermo Scientific HAAKE Instruction Manual

Table of Contents

Advertisement

Maintenance by the User

12. Maintenance by the User
21
16
13
22
18
17
20
9
8
1
12
4
11
4
8
7
17
18
14
21
12.1 Exchange of the measuring tube and the jacket
tube
Unscrew the screws 4 and remove brace 11.
1
Loosen threaded bush 18 with socket spanner
2
no. 003-2110.
Unscrew connecting rods 20.
3
Now remove both cover lids 7 and 9.
4
Exchange jacket tube 12.
5
In order to exchange the measuring tube remove
6
bush 19 and gasket 17.
Assembly
Screw jacket tube 12, flat gasket 8, cover 7 and
7
9 together with the connecting rods 20.
!
Don't tighten the connecting rods initially.
Insert measuring tube 1 with gasket 17 into the
8
covers 7 and 9.
Mount brace 11, don't tighten the screws 4
9
initially.
Please tighten the connecting rods 20 evenly and
10
carefully.
!
Recommended tightening moment 0,8-1,0 Nm.
Please control, whether the jacket tube is
moveable axially.
Fix the screws 4.
11
Please control, whether the jacket tube is move-
able axially.
Fix threaded bush 18 with the socket spanner
12
003-2110 slightly.
20