2. Run the IPSetup.exe utility from any IBM®-compatible PC connected
to the same network as the Bridge. The utility automatically searches
the network and reports any bridges found.
3. If more than one bridge is installed on the network, each can be
identified by its MAC address. The MAC address is recorded on the
label near the RJ45 connector on the Bridge board.
4. Record the IP address assigned to the bridge. This is the IP address
required for configuring a workstation or DCS to query the SOLA II
Flare. On the TCP/IP Modbus frame, the unit ID must always be set to
1.
Figure C–2. Example display with DHCP server present
To set a static IP, enter the IP address and press the Set--> button.
The SOLA II Flare offers the option for the user to remotely connect to the
SOLA II Flare via web browser and be presented with a full representation of
the display on the front of the unit. This provides the user with all of the same
functionality as standing in front of the unit itself.
Figure C–3. Browser connect screen
Appendix C
TCP/IP Bridge
SOLA II Flare User Manual