The Rotary Valve
The Mixing Chamber
Thermo Scientific
6. Connect the flow meter to the Meas. Clean Air 1 port and adjust the
Clean Air 1 regulator to obtain the required flow.
7. Turn the Meas. Clean Air 1 three-way valves to point up for normal
operation.
8. Turn the Meas. Clean Air 2 three-way valves to point down for flow
measurement.
9. Connect the flow meter to the Meas. Clean Air 2 port and adjust the
Clean Air 2 regulator to obtain the required flow.
10. Turn the Meas. Clean Air 2 three-way valves to point up for normal
operation.
11. Turn the Meas. Clean Air 3 three-way valves to point down for flow
measurement.
12. Connect the flow meter to the Meas. Clean Air 3 port and adjust the
Clean Air 3 regulator to obtain the required flow.
13. Turn the Meas. Clean Air 3 three-way valves to point up for normal
operation.
14. Restart the analyzer according to
For detailed instructions on maintaining, removing, and replacing the rotary
valve, refer to
Appendix
To remove the mixing chamber, follow the steps below.
1. Shut down the system according to
2. Loosen the fittings that connect the tubing to the mixing chamber.
3. Loosen the screw in the center of the bracket holding the mixing
chamber to the bottom of the enclosure.
4. Carefully slide the mixing chamber from the bracket.
To reinstall the mixing chamber, follow the steps below.
1. Carefully slide the mixing chamber into the bracket.
2. Tighten the screw in the center of the bracket to hold the mixing
chamber to the bottom of the enclosure.
Chapter
B.
Chapter
Maintenance & Troubleshooting
4.
3.
SOLA II Flare User Manual
Chapter 8
8-7