18. Insert the ends of the heater wires into the terminal block and tighten
them. Bend the wires as shown in the figure below.
Figure 8-12. Heater wires inserted into terminal block
19. Inspect the O-ring around the open end of the pyrolyzer housing. If
necessary, replace it and apply a lubricant compound, such as Dow
Corning® 111, around it.
Figure 8-13. Pyrolyzer housing O-ring
20. Gently slide the lid into the housing.
Note The pyrolyzer tube, insulation, and Kaowool insulation should still be
attached to the lid. Slide the lid back in carefully to avoid damaging the
pyrolyzer tube. ▲
21. Use the six screws and washers removed earlier to secure the lid to the
housing.
Maintenance & Troubleshooting
SOLA II Flare User Manual
Chapter 8
8-14