Thermo Scientific SOLA II Flare User Manual page 91

Sulfur online analyzer
Table of Contents

Advertisement

Thermo Scientific
9. As shown in the following figure, the heater wires are attached to the
terminal block. Remove the heater wires from the terminal block and
straighten the wires. Remove the fiberglass insulation that covers the
heater wires. Set the insulation aside. You will need it later.
Figure 8-5. Heater wires and thermocouple
Figure 8-6. Remove the fiberglass insulation
10. Loosen the fittings around the heater wires and remove the wire sleeves.
Set the sleeves and fittings aside.
11. Loosen the thermocouple fitting and remove the thermocouple from
the housing. Set it aside for re-installation later.
12. You should now be able to remove the heater from the housing.
Maintenance & Troubleshooting
SOLA II Flare User Manual
Chapter 8
8-11

Advertisement

Table of Contents
loading

Table of Contents