Thermo Scientific SOLA II Flare User Manual page 112

Sulfur online analyzer
Table of Contents

Advertisement

Parameter
Display
Software
Control Mode
Autotune Set Point
Autotune
Closed-Loop Set Point
Idle Set Point
Heat Proportional Band
Heat Hysteresis
Cool Proportional Band
Cool Hysteresis
Time Integral
Time Derivative
Dead Band
Open Loop Set Point
Parameter
Display
Software
Alarm Low Set Point
Alarm High Set Point
Thermo Scientific
Table A–11. Pyrolyzer settings: Operations Page > Control Loop 1
Default
Display
Table A–12. Pyrolyzer settings: Operations Page > Alarm 1
Default
Display
Software
Auto
90%
No
1100°C
24°C
14°C
2°C
14°C
2°C
180
0
0°C
0.0%
Software
-5°C
5°C
SOLA II Flare User Manual
Appendix A
Watlow EZ Zone PM6 Series
Comment
The loop will use automatic control.
The percentange of the current set
point that will be used by autotune.
The set point the controller will
automatically control to.
The closed loop set point that can be
triggered by an event state.
The PID proportional band for the heat
outputs.
The process value needs to be this far
into the On region before the output
turns on.
The PID proportional band for the cool
outputs.
The process value needs to be this far
into the On region before the output
turns on.
The PID integral (seconds).
The PID derivative (seconds).
The offset to the proportional band.
The fixed level of output power when
in manual (open loop) mode.
Comment
The set point that will trigger a low
alarm.
The set point that will trigger a high
alarm.
A-8