Post Calibration - Setup at the Job Site ............5-1 General ....................5-1 Procedure .................... 5-1 Chapter 6 The Current Output .................... 6-1 Hold the Current Output ..............6-1 Clear All Holds ................... 6-2 Check the Input Signal ............... 6-2 Thermo Fisher Scientific SGD-O User Guide...
Page 10
Appendix D Gauge Calibration Records ................D-1 Appendix E Commonly Used Parameter & Command DACs .......... E-1 Toxic & Hazardous Substances Tables ............F-1 Appendix F Appendix G Related Technical Bulletins ................G-1 viii SGD-O User Guide Thermo Fisher Scientific...
The triangular icon displayed with a warning may contain a lightning bolt or the radiation symbol, depending on the type of hazard. Thermo Fisher Scientific SGD-O User Guide...
Page 12
Note Notes emphasize important or essential information or a statement of company policy regarding an operating procedure, practice, condition, etc. SGD-O User Guide Thermo Fisher Scientific...
Chapter 1 Product Overview Introduction The Thermo Scientific SGD-O system consists of a radiation source in a lead-filled housing, an ion chamber detector with an amplifier, and the SGD-O transmitter. The transmitter processes the detector signal to provide a signal output proportional to the density of the slurry, the mass of solids added per volume of carrier, or both.
Page 14
Product Overview Introduction Figure 1–2. Example of a shutterless unit installed on a high pressure pipe Figure 1–3. Example of a unit with shutter (low pressure application) SGD-O User Guide Thermo Fisher Scientific...
Regulations for the NRC, U.S. Department of Transportation, and the Canadian Nuclear Safety Commission are continuously being updated; contact Thermo Fisher Scientific for information on proper source disposal. Thermo Fisher Scientific SGD-O User Guide...
The Transmitter The SGD-O transmitter is a microprocessor-based system that processes the signal from the detector (0–10 Vdc) to provide an output signal (4–20 mA) representing the amount of solids (proppant) added per volume unit of carrier and/or the bulk density of the slurry.
Two optional relay outputs can be provided on each I/O board. Note If the detector will be connected directly to customer equipment (i.e. the SGD-O transmitter will not be used), the detector output is 0 to 10 Vdc. Thermo Fisher Scientific...
Required In addition to this guide, the following documents must be read and understood by all persons installing, using, or maintaining this equipment: Documentation SGD-O Installation Guide (p/n 717913) Gamma Radiation Safety Guide (p/n 717904) SGD-O User Guide...
Chapter 2 Getting Started Powering the There are three input power options for the SGD-O transmitter: Transmitter 12 Vdc version (standard) ● 24 Vdc (optional) ● 115/230 Vac (optional) ● The transmitter power cable is configured to operate the 12 Vdc version only.
In this example, force the current output to 20 mA. The command code is 5. 1. Press: EXIT SETUP, 5, down arrow. 2. The display shows two options. Press the left arrow to reject the command or the right arrow to execute it. SGD-O User Guide Thermo Fisher Scientific...
± 0.5 Vdc of specified voltage. If one or both of the voltages are incorrect or not there at all, either the transmitter to detector cable is bad or the transmitter is not outputting the ±15 Vdc properly. Thermo Fisher Scientific SGD-O User Guide...
Page 22
Note If the detector output is above 8.0 Vdc, the pipe ID is 6 inches or smaller, and the measurement range is under 10 PPA, it is acceptable to use the voltage and continue to the calibration procedure in Chapter SGD-O User Guide Thermo Fisher Scientific...
For optimum performance, calibrate the gauge exactly as described below. It is assumed that the gauge has been installed properly, the shutter is open (if applicable), the pipe is empty of any fluid, and that the SGD-O transmitter is being used.
Page 24
9. Reset the CAL/REF ratio to the default value of 1.000 (DAC 086003). 10. Standardize the gauge on water (DAC 121). This will set the standardization value (water value) that will be used to calculate the ZnCl calibration value. SGD-O User Guide Thermo Fisher Scientific...
Page 25
16. Recall the density slope correction factor (DAC 049003). Record this value. 17. Recall the carrier specific gravity (DAC 085003). Record this value. Note The values recorded in steps 16 and 17 are required for troubleshooting purposes. Keep these values handy. Thermo Fisher Scientific SGD-O User Guide...
Page 26
20. The displayed density should then be 1.000 g/ml. The gauge is now calibrated for use of water as the carrier. Note This calibration will be valid as long as the detector is not moved or removed from the housing. SGD-O User Guide Thermo Fisher Scientific...
Chapter 4. Procedure Once the SGD-O gauge is in service, one more operation is necessary for proper indication of the Pounds of Sand Added and the Pounds per Gallon indications of the gauge. Note Reference Chapter 2 for instructions on how to use a DAC.
Page 28
Sand Added and the Pounds per Gallon indications will be correct. When the job is finished, return the carrier density to its original value (DAC 085003). To prepare for the next job, perform a standardization on a pipe full of clean water (DAC 121). SGD-O User Guide Thermo Fisher Scientific...
Note The initial value in this formula is the measured value of the 20 mA output when the correction factor is equal to 1.0. 3. Hold the current output at 12 mA (DAC 2). Thermo Fisher Scientific SGD-O User Guide...
The output of the detector (input to the transmitter) is a voltage signal 0–10 Vdc maximum. Read the signal (DAC 112003). The value is Signal displayed in scientific notation and represents a range of 0 to 10 volts with the numbers 0 to 6.554E4 (65540). SGD-O User Guide Thermo Fisher Scientific...
Adjusting the Gain Settings Overview Users of the SGD-O density gauge often move the detector assembly from application to application. Among these different applications are variations in sizes and schedules of pipes, as well as activity and ages of radioactive sources.
Page 32
A 220 pF capacitor will make the gain higher and a 470 pF capacitor will make the gain lower. If you are having trouble getting the gain either high enough or low enough, check the value of C13. SGD-O User Guide Thermo Fisher Scientific...
220, 330, or 470 pF 2X or 3X Gain Control Figure 7–2 shows how the jumpers and resistors work together to control the gain of the AR1 op-amp. Settings Figure 7–2. Gain control of the AR1 op-amp Thermo Fisher Scientific SGD-O User Guide...
Page 34
W2 jumper + W3 jumper + This leaves the W5 leg at 49.9K and puts R36 R36 changed to 49.9 kohm and R37 in parallel, changing the current to the op-amp. SGD-O User Guide Thermo Fisher Scientific...
Page 35
R36 will change R36 to 4.12 kohm. Everything else remains at default values. W2 jumper + W3 jumper Everything else remains at default values. Figure 7–3. C13 in the control circuit of the FSU detector Thermo Fisher Scientific SGD-O User Guide...
Chapter 8 Maintenance Overview The SGD-O is fairly maintenance free, but you should follow the schedule below to remain in compliance. The tasks are described in further detail in the sections that follow. Table 8–1. Maintenance schedule Task Interval Complete a shutter check (if the system Every 6 months.
Page 38
Figure 8–2. DOT 7A Type A label with General License information The DOT 7A Type A label example above (p/n 702272) provides General License information and is only used on instruments going to customers with a General License. SGD-O User Guide Thermo Fisher Scientific...
Page 39
License information and is only used on instruments going to customers with a Specific License. Figure 8–4. Caution tag example Note None of these tags should be removed, painted over, or remain illegible. Thermo Fisher Scientific SGD-O User Guide...
Thermo Fisher. If you purchased a gauge prior to the above date, you can request to be added to Thermo Fisher’s leak test service. SGD-O User Guide Thermo Fisher Scientific...
Memory memory. Note This procedure assumes that the SGD-O gauge was fully calibrated as described in this manual and that the water standardization value and slope correction value were recorded from the last calibration. If these two values are not available, it is not possible to perform this procedure.
Open the enclosure and verify that the keypad cable is properly installed on the CPU board. If transmitter still does not display, the CPU board may be malfunctioning. Refer to the SGD-O installation manual for instructions on replacing boards in the transmitter. The DC Power Fuse If you suspect a problem with the DC power fuse, power down the unit and measure the resistance with a multimeter.
“The FSU Pre-Amp Board.” The locations of the components and jumpers referenced in this procedure are shown in the figures below. Figure 9–2. SGD-O FSU pre-amp board Figure 9–3. Power supply board Thermo Fisher Scientific SGD-O User Guide...
Page 44
Move the W1 jumper to W2 (keeping in mind that the center hole is the common connection). In most cases, you should now be able to use the R35 pot to adjust the output to 9.6 Vdc on air. SGD-O User Guide Thermo Fisher Scientific...
Measure the Vdc on test point 1 and use the R1 pot to set this voltage to +5 mV. The FSU Pre - This section provides a procedure for troubleshooting the SGD-O FSU pre-amp board (p/n 885623). The procedure will aid technical personnel in Amp Board...
Page 46
Measure +15V and -15V supply current tap voltages: +15V current: Voltage drop across R1 should be between 130 mV and 280 mV. -15V current: Voltage drop across R3 should be between 130 mV and 280 mV. SGD-O User Guide Thermo Fisher Scientific...
T2 (at TP3): Should be between (T1+14 ms) and (T1+37 ms). T3 (at TP2): Should be between (T1+14 ms) and (T1+37 ms) (this is a negative pulse). Remove the 4.7 x 10 ohm resistor. Thermo Fisher Scientific SGD-O User Guide...
Additionally, the Type A shipping certificate is issued to the instrument as a single component consisting of the detector, source housing, and pipe. If the source housing is separated from the pipe, it is no longer covered by the Type A shipping certificate. SGD-O User Guide Thermo Fisher Scientific...
Troubleshooting & Service Warranty Warranty Thermo Scientific products are warranted to be free from defects in material and workmanship at the time of shipment and for one year thereafter. Any claimed defects in Thermo Scientific products must be reported within the warranty period. Thermo Fisher shall have the right to inspect such products at Buyer’s plant or to require Buyer to return such...
I/O board, 24 Vdc input 886675-3 I/O board, 12 Vdc input 886675-5 I/O board, no DC input CP6879901 VPI board, modified for SGD-O Table A–2. Ion Chamber Detector Description 050273 Complete ion chamber detector, with boards 050274 Electronics board set...
Page 52
Table A–5. Spares Description 050268 Housing with detector, no lid 45373B Plastic spacer (ion chamber support) 542630 O-ring for lid 54860Z Rubber grommet (Elast-o-vier) 885745 Detector housing with support 885746 Insulation kit, with desiccant packs SGD-O User Guide Thermo Fisher Scientific...
-58°F to +140°F (-50°C to +60°C) Housing Source head: Model 5190/5192 Detector: Stainless steel 316 or carbon steel Table B–2. SGD-O transmitter specifications Input Standard DC power: 12 Vdc (9–15 V), 15 W Optional DC power: 24 Vdc (20–28 V), 12 W Optional AC power: 115/230 Vac (100–240 V), 50/60 Hz, 25 VA...
Page 54
Specifications Keypad 10 keys, external Housing Stainless steel 316 Mounting Transmitter can be mounted 150 ft. (45.72 m) from the detector SGD-O User Guide Thermo Fisher Scientific...
Record the carrier specific gravity. 085003 Displayed density should be 1.812 g/ml. Adjust slope 049003 correction factor if necessary until density is 1.812 g/ml. Clear all holds. Displayed density should be 1.000 g/ml. (Record value.) Completed by: Thermo Fisher Scientific SGD-O User Guide...
0.9982 g/cm Carrier attenuation coefficient 019003 0.086 cm Solids attenuation coefficient 020003 0.077 cm Weeks since last reference 054003 Enable less than 0 and greater than 043030 100% Enable phantom measurements (HV 045020 control) Thermo Fisher Scientific SGD-O User Guide...
Page 68
(LevelPRO) High voltage hold value 070003 Table E–2. Common command DACs Command Description Preload configuration for SGD-O oilfield in metric units 1014 Preload configuration for SGD-O oilfield in English units 1016 Hold input at value indicated by parameter 133003 Hold current output at 50% (12 mA for a 4–20 mA) Hold current output at maximum (20 mA for 4–20 mA)
Appendix F Toxic & Hazardous Substances Tables The English and Chinese versions of the Toxic and Hazardous Substances tables are below. Thermo Fisher Scientific SGD-O User Guide...
This appendix provides a list of technical bulletins related to this product as of the release date of this document revision. For the most current bulletins, please go to thermoscientific.com. Enter the product name (SGD-O) as the search term and click the Resources tab. Table G–1. Technical bulletins TB #...