7. Remove the six screws and washers from the lid of the housing. Set
them aside for later use.
Figure 8-3. Remove the pyrolyzer housing lid
8. Gently lift the lid off and set it aside.
Note The pyrolyzer tube, insulation, and Kaowool™ insulation will be attached
to the lid. Carefully lift all of it out to avoid damaging the pyrolyzer tube. ▲
Figure 8-4. Housing lid with insulation and pyrolyzer tube attached
Maintenance & Troubleshooting
SOLA II Flare User Manual
Chapter 8