Thermo Scientific SOLA II Flare User Manual page 108

Sulfur online analyzer
Table of Contents

Advertisement

Parameter
Display
Software
Heat Algorithm
Cool Algorithm
TRU-TUNE+ Enable
Autotune Aggressiveness
User Failure Action
Input Error Failure
Open Loop Detect Enable
Ramp Action
Low Set Point
High Set Point
Set Point Open Limit Low
Set Point Open Limit High
Parameter
Display
Software
Output Function
Output Control
Output Time Base
Output Low Power Scale
Output High Power Scale
Thermo Scientific
Table A–3. Pyrolyzer settings: Setup Page > Control Loop 1
Default
Display
Table A–4. Pyrolyzer settings: Setup Page > Output 1
Default
Display
Software
PID
Off
No
Critical
User
Off
No
Off
0°C
1150°C
-100.0%
100.0%
Software
Heat
Fixed Time Base
0.5
0%
15%
SOLA II Flare User Manual
Appendix A
Watlow EZ Zone PM6 Series
Comment
The method of heat control.
Critical damped aggressiveness of
autotuning calculations.
When user switches to manual mode,
controller sets output power to last
open-loop set point entered by user.
When an input error switches control
to manual mode, controller sets output
to 0%.
The minimum value of the closed loop
set point range.
The maximum value of the closed loop
set point range.
The minimum value of the open loop
set point range.
The maximum value of the open loop
set point range.
Comment
The power output will never be less
than the value specified and will
represent the value at which output
scaling begins.
The power output will never be greater
than the value specified and will
represent the value at which output
scaling stops.
A-4