7.6

Analog outputs

The analog outputs (optional extra) can each be configured as current or voltage outputs (current signal,
voltage signal) and are freely scalable.
Device dialog box
Parameter
Parameter
Channel designa-
tion
Channel descrip-
tion
Signal source
Output signal
Scaling
start
?
Scale
end
?
Selection/settings
AO ...
Use default text or enter other text.
Analog output ...
Use default text or enter other text.
Analog selector
No selection
Physical output signal
Voltage 0 to 10 V
Current 0 to 20 mA
Current 4 to 20 mA
Voltage 10 to 0 V
Current 20 to 0 mA
Current 20 to 4 mA
-99999 to 0 to +99999
-99999 to 100 to +99999
7 Configuration
Description
Name (abbreviation) with max. 5 charac-
ters that is used in the visualizations.
Description (designation) with max. 21
characters
Analog signal that is issued via the ana-
log output.
In the event of "No selection" a voltage of
0 V is output.
Voltage signal
Current signal
Current signal
Voltage signal, inverse
Current signal, inverse
Current signal, inverse
Start value of value range
End value of value range
111