JUMO LOGOSCREEN 601 Operating Manual page 159

Paperless recorder with touchscreen
Hide thumbs Also See for LOGOSCREEN 601:
Table of Contents

Advertisement

Parameter
Upper switching
differential
Lower switching
differential
Alarm text upper
limit
Alarm text lower
limit
Show auxiliary
line 1, 2
Auxiliary line val-
ue 1, 2
Auxiliary line col-
or 1, 2
Input signal
Channel 1 has a special task; it is used as the reference curve for the other channels in the group. All
other channels in the group can be compared with channel 1 independently of one another. If the con-
figured tolerance band is exited, an alarm occurs, and an alarm text is displayed in the "status and title
line" and entered in the alarm list and event list.
NOTE!
The tolerance band comparison is only possible within the scaling limits. If a channel drops below or ex-
ceeds the range, this will result in both a min. and a max. alarm as with all other functions (in this case,
pos. tolerance and neg. tolerance).
Selection/settings
Only available for channel 1 of a group.
-100 to -2 to 0
Only available for channel 1 of a group.
0 to 2 to 100
Available as from channel 2 of a group.
Use default text or enter other text.
Available as from channel 2 of a group.
Use default text or enter other text.
Two auxiliary lines (level lines) can be shown for each channel.
No, Yes
The two auxiliary lines can be used, for example, to define a lower and upper limit.
As a result, the user can see in the diagram whether the current value is within the
defined range.
0 to 100
The default setting for the auxiliary line color corresponds to the color of the analog
channel concerned (Configuration > Display > Colors).
Use default color or select another color. Auxiliary line color
7 Configuration
Description
If there is a positive tolerance violation,
the current measured value from chan-
nel 2 to 6 must first drop below the cur-
rent measured value of channel 1, plus
the upper tolerance band limit and upper
switching differential, in order for the
alarm to be canceled.
If there is a negative tolerance violation,
the current measured values from chan-
nel 2 to 6 must first exceed the current
measured value of channel 1, plus the
lower tolerance band limit and lower
switching differential, in order for the
alarm to be canceled.
The text entered (max. 31 characters) is
displayed and entered in the alarm and
event list if the selected channel exits the
tolerance in positive direction.
The text entered (max. 31 characters) is
displayed and entered in the alarm and
event list if the selected channel exits the
tolerance in negative direction.
If "Yes" is selected, the auxiliary line is
shown in the vertical and horizontal
curve diagram if this has been activated
generally in the configuration of the dia-
gram view.
Auxiliary line value
159