JUMO LOGOSCREEN 601 Operating Manual page 266

Paperless recorder with touchscreen
Hide thumbs Also See for LOGOSCREEN 601:
Table of Contents

Advertisement

11 Annex
RTD temperature probe
Designation
Standard
Pt50
DIN EN 60751:2009
IEC 60751:2008
Pt100
DIN EN 60751:2009
IEC 60751:2008
Pt500
DIN EN 60751:2009
IEC 60751:2008
Pt1000
DIN EN 60751:2009
IEC 60751:2008
Pt100
JIS C 1604:1981
Pt50
GOST 6651-2009 A.2
Pt100
GOST 6651-2009 A.2
Cu50
GOST 6651-2009 A.3
Cu100
GOST 6651-2009 A.3
Ni100
DIN 43760 (1987)
Ni100
GOST 6651-2009 A.5
Connection type
Ambient temperature influence
Sensor line resistance
a
Accuracy refers to the measuring range.
Resistance transmitter and resistance/potentiometer
Designation
Resistance transmitter
Resistance/potentiometer
Ambient temperature influence
Connection type
Resistance transmitter
Resistance/potentiometer
Smallest measuring span
Sensor line resistance
Resistance values
a
Accuracy refers to the maximum measuring range. Small measuring spans lead to reduced linearization accuracy.
266
ITS
ITS-90
ITS-90
ITS-90
ITS-90
IPTS-68
ITS-90
ITS-90
ITS-90
ITS-90
IPTS-68
ITS-90
2/3/4-wire
≤ 50 ppm/K
Max. 10 Ω per cable for two-wire circuit
Max. 30 Ω per cable for three/four-wire circuit
Measuring range
0 to 4000 Ω
0 to 400 Ω
0 to 4000 Ω
≤ 100 ppm/K
Three-wire circuit
Two/three/four-wire circuit
60 Ω
Max. 10 Ω per cable for two-wire and three-wire circuits
Freely programmable within the limits, in increments of 0.1 Ω
Measuring range
-200 to +850 °C
-200 to +850 °C
-200 to +850 °C
-200 to +850 °C
-200 to +649 °C
-200 to +850 °C
-200 to +850 °C
-180 to +200 °C
-180 to +200 °C
-60 to +250 °C
-60 to +180 °C
a
Accuracy
≤ 0.1 %
≤ 0.1 %
≤ 0.1 %
a
Accuracy
Measur-
ing cur-
rent
≤ 0.1 %
500 μA
≤ 0.1 %
500 μA
≤ 0.1 %
50 μA
≤ 0.1 %
50 μA
≤ 0.1 %
500 μA
≤ 0.1 %
500 μA
≤ 0.1 %
500 μA
≤ 0.4 %
500 μA
≤ 0.4 %
500 μA
≤ 0.2 %
500 μA
≤ 0.2 %
500 μA
Measuring current
50 μA
500 μA
50 μA

Advertisement