JUMO flowTRANS MAG S01 Operating Manual

JUMO flowTRANS MAG S01 Operating Manual

Electromagnetic flowmeter for the process industry and hygienic applications
Hide thumbs Also See for flowTRANS MAG S01:
Table of Contents

Advertisement

JUMO flowTRANS MAG S/H
Electromagnetic flowmeter
for the process industry and hygienic applications
Operating Manual
for device versions 406012 - 406019
40601200T90Z001K000
V5.00/EN/00643278

Advertisement

Table of Contents
loading

Summary of Contents for JUMO flowTRANS MAG S01

  • Page 1 JUMO flowTRANS MAG S/H Electromagnetic flowmeter for the process industry and hygienic applications Operating Manual for device versions 406012 - 406019 40601200T90Z001K000 V5.00/EN/00643278...
  • Page 3: Table Of Contents

    JUMO flowTRANS MAG S/H - remote mount design ....... .
  • Page 4 4.2.2 JUMO flowTRANS MAG H - Variable process connections ......40 Mounting ..............41 4.3.1...
  • Page 5 Contents 5.4.3 Connection via cable conduits ..........58 5.4.4 Potting the terminal box .
  • Page 6 Alarm Simulation............147 Filling operation - JUMO flowTRANS MAG S02/H02 ....... 149 8.6.1...
  • Page 7 Sensor JUMO flowTRANS MAG S01/02 ........
  • Page 8 Sensor JUMO flowTRANS MAG H01/02 ........
  • Page 9: Safety

    The symbol on the nameplate refers to: Other applicable device documentation This document "JUMO flowTRANS MAG S/H - Operating Manual for Devices 406012 - 406019" is sup- plemented by the documents listed below: Document name...
  • Page 10: Intended Use

    1 Safety Intended use This device serves the following purposes: • To transfer liquid, pulpy or pasty measurement media with electrical conductivity • To measure the flow of the operating volume or measurement units (with constant pressure/tempera- ture) if a physical measuring unit has been selected Intended use also includes the following points: •...
  • Page 11: Symbols And Signal Words

    1 Safety Symbols and signal words DANGER – Serious damage to health/risk to life! This symbol in connection with the signal word "Danger" indicates an imminent threat of danger. Failure to observe the safety information results in death or serious injuries. DANGER –...
  • Page 12: Nameplate

    Nameplate IMPORTANT (NOTE)! There is the additional document "JUMO flowTRANS MAG S/H - Safety Manual Ex for devices 406012 - 406019" for the measuring systems that are used in potentially explosive areas. The information and data contained in the document must also be strictly observed.
  • Page 13 1 Safety Model number (for technical version details, see data sheet or order confirmation) Serial number for identification by the manufacturer Nominal width and nominal pressure levels Material: flange/lining/electrode = Maximum admissible measurement medium temperature = Maximum admissible ambient temperature Calibration value Qmax DN Calibration value Ss (range) Calibration value Sz (zero point)
  • Page 14 1 Safety Nameplate Model number (for technical version details, see data sheet or order confirmation) Serial number for identification by the manufacturer Nominal width and nominal pressure levels = Maximum admissible measurement medium temperature = Maximum admissible ambient temperature Protection type according to DIN EN 60529 Calibration value Qmax DN Transmitter communication protocol Software version...
  • Page 15 1 Safety Model number (for technical version details, see data sheet or order confirmation) Serial number for identification by the manufacturer Nominal width and nominal pressure levels Material: flange/lining/electrode Customer-specific TAG number (if provided) = Maximum admissible measurement medium temperature = Maximum admissible ambient temperature Protection type according to DIN EN 60529 Calibration value Qmax DN...
  • Page 16 1 Safety Model number (for technical version details, see data sheet or order confirmation) Serial number for identification by the manufacturer Customer-specific TAG number (if provided) = Maximum admissible ambient temperature Protection type according to DIN EN 60529 Voltage supply Transmitter communication protocol Software version Order number...
  • Page 17: Transport Safety Information

    1 Safety Nameplate Model number (for technical version details, see data sheet or order confirmation) Serial number for identification by the manufacturer Voltage supply = Maximum admissible ambient temperature Transmitter communication protocol CE mark Order number Customer-specific TAG number (if provided) Protection type according to DIN EN 60529 (10) Software version...
  • Page 18: Safety Information For The Electrical Installation

    1 Safety 1.10 Safety information for the electrical installation The electrical connection must only be established by authorized qualified personnel according to the electrical diagrams. Observe the notes in the manual concerning the electrical connection. Otherwise, the electrical protec- tion rating may be adversely affected. The flow measuring system and the transmitter housing must be grounded.
  • Page 19: Technical Limit Values

    1 Safety 1.12 Technical limit values The device is only designed for use within the technical limit values specified on the nameplate and in the data sheets. The following technical limit values must be observed: • The admissible operating pressure (PS) and the admissible temperature (TS) must not exceed the pressure temperature ratings (p/T ratings).
  • Page 20: Safety Information For Inspections And Maintenance

    These measures must be confirmed in writing in the declaration of decontamination. Return address and documents IMPORTANT (NOTE)! All documents important for the return as well as the return address of the manufacturer are available at http://reparaturdienst.jumo.info.
  • Page 21: Integrated Management System

    1 Safety 1.16 Integrated management system The manufacturer has an integrated management system consisting of: • Quality management system ISO 9001:2008 • Environmental management system ISO 14001:2004 + Cor1:2009 • Energy management system ISO 50001:2011 This environmental approach is part of our corporate policy. The impact to the environment and humans is to be kept as low as possible during the manufacturing, storage, transport, use, and disposal of our products and solutions.
  • Page 22 1 Safety...
  • Page 23: Design And Function

    Design and function 2 Design and function Measuring principle The Faraday Laws of Induction provide the basis for the electromagnetic flowmeter measurement. A voltage is generated in a conductor as it moves through a magnetic field. When this measuring principle is applied through a device, the conductive measurement medium flows through a pipe in which a magnetic field is generated perpendicular to the flow direction (see figure).
  • Page 24: Device Versions

    Device designation IMPORTANT (NOTE)! There is the additional document "JUMO flowTRANS MAG S/H - Safety Manual Ex for devices 406012 - 406019" for the measuring systems that are used in potentially explosive areas. The information and data contained in the document must also be strictly observed.
  • Page 25: Functionalities

    2 Design and function 2.3.2 Functionalities JUMO flowTRANS MAG electromagnetic flowmeters are available with basic and advanced functions. Model Type series Functionality 406012 JUMO flowTRANS MAG S01 Basic functions 406015 JUMO flowTRANS MAG H01 406013 JUMO flowTRANS MAG S02 Basic functions and advanced functions...
  • Page 26: Design

    (see chapter 2.4.3 "Housing design type", page 27) should also be considered. See also chapter 6.2 "Notes on combining the sensor and transmitter," page 47 in the "JUMO flowTRANS MAG S/H Safety manual Ex for the devices 406012 - 406019."...
  • Page 27: Housing Design Type

    2 Design and function 2.4.3 Housing design type JUMO flowTRANS MAG electromagnetic flowmeters are deployed in a wide variety of areas of applica- tion. For devices with a remote mount design type, this distinction only relates to the transmitter housing. Single-compartment housing For single-compartment housing, the electronics compartment and the terminal compartment are not separated from each other in the transmitter.
  • Page 28: Overview Of Device Versions

    2 Design and function Overview of device versions 2.6.1 JUMO flowTRANS MAG S/H - compact design JUMO flowTRANS MAG S01/S02 406012/1-0 (without Ex-protec- 406012/1-1 (Ex-protection zone 2) 406012/1-1 (Ex-protection zone 1) tion) 406013/1-0 (without Ex-protec- 406013/1-1 (Ex-protection zone 2) 406013/1-1 (Ex-protection zone 1)
  • Page 29: Jumo Flowtrans Mag S/H - Remote Mount Design

    2 Design and function 2.6.2 JUMO flowTRANS MAG S/H - remote mount design JUMO flowTRANS MAG S01/S02 406012/2-0 (without Ex-protec- 406012/2-1 (Ex-protection zone 2) 406012/2-1 (Ex-protection zone 1) tion) 406013/2-0 (without Ex-protec- 406013/2-1 (Ex-protection zone 2) 406013/2-1 (Ex-protection zone 1)
  • Page 30 2 Design and function JUMO flowTRANS MAG H01/H02 406015/2-0 (without Ex-protection) 406015/2-1 (Ex-protection zone 2) 406016/2-0 (without Ex-protection) 406016/2-1 (Ex-protection zone 2) Sensor Sensor ATEX/IEC Gas zone 2 Dust zone 21, 22 JUMO flowTRANS MAG 01/02 406018/2-0 (without Ex-protection) 406018/2-1 (Ex-protection zone 2)
  • Page 31: Transport And Storage

    Transport and storage 3 Transport and storage Checking the delivery After unpacking the devices, check them immediately for any possible damage that may have occurred due to improper transport. Transport damage must be documented on the shipment documents. All claims for damages must be submitted to the shipper immediately and before the installation. Transporting flanged devices 3.2.1 Flange devices up to DN 400...
  • Page 32: Flange Devices Greater Than Dn 400

    3 Transport and storage 3.2.2 Flange devices greater than DN 400 WARNING – Risk of injury! Do not lift flange devices directly on the connection box or housing centre! CAUTION – Device damage! • Do not use chains to lift the flange devices! •...
  • Page 33: Mounting

    4 Mounting IMPORTANT (NOTE)! There is the additional document "JUMO flowTRANS MAG S/H - Safety Manual Ex for devices 406012 - 406019" for the measuring systems that are used in potentially explosive areas. The information and data contained in the document must also be strictly observed.
  • Page 34: Devices With A Wafer Design

    4 Mounting 4.1.2 Devices with a wafer design For mounting devices with a wafer design, the manufacturer offers a mounting set as an accessory con- sisting of threaded rods, nuts, washers and centering sleeves. G01348 Threaded rod Nut with washer Centering sleeves 4.1.3 Installing the measuring pipes...
  • Page 35: Tightening Torques

    4 Mounting Tightening torques 4.2.1 JUMO flowTRANS MAG S/H - Flanged/connection flange design IMPORTANT (NOTE)! The specified torques only apply to greased threads and to pipelines that are not under tension. Nominal Nominal Maximum tightening torque (Nm) with lining width...
  • Page 36 4 Mounting Nominal Nominal Maximum tightening torque (Nm) with lining width pressure Hard rubber, soft rubber PTFE, PFA, ETFE Ceramic carbide [mm (inch)] levels Flange material Flange material Flange material Steel Stainless Steel Stainless Steel Stainless Steel Steel Steel DN 40 (1 1/2") PN 40 30.44 23.71 56.06...
  • Page 37 4 Mounting Nominal Nominal Maximum tightening torque (Nm) with lining width pressure Hard rubber, soft rubber PTFE, PFA, ETFE Ceramic carbide [mm (inch)] levels Flange material Flange material Flange material Steel Stainless Steel Stainless Steel Stainless Steel Steel Steel DN 150 (6") PN 16 46.99 23.7...
  • Page 38 4 Mounting Nominal Nominal Maximum tightening torque (Nm) with lining width pressure Hard rubber, soft rubber PTFE, PFA, ETFE Ceramic carbide [mm (inch)] levels Flange material Flange material Flange material Steel Stainless Steel Stainless Steel Stainless Steel Steel Steel DN 500 (20") PN 10 141.6 101.4...
  • Page 39 4 Mounting Nominal Nominal Maximum tightening torque (Nm) with lining width pressure Hard rubber, soft rubber PTFE, PFA, ETFE Ceramic carbide [mm (inch)] levels Flange material Flange material Flange material Steel Stainless Steel Stainless Steel Stainless Steel Steel Steel DN 1000 (40") PN 10 541.3 419.2 Upon re-...
  • Page 40: Jumo Flowtrans Mag H - Variable Process Connections

    PN 16 2860 Upon re- quest 4.2.2 JUMO flowTRANS MAG H - Variable process connections Nominal width in mm (inch) Maximum tightening torque in Nm DN 1 to 2 (1/25 to 3/32) 0.2 (for PVC/POM) 3 (for 1.4571) DN 3 to 10 (3/8)
  • Page 41: Mounting

    4 Mounting Mounting 4.3.1 Flow direction The device records the flow in both directions. Default setting: forward flow direction Identification marking: • Arrow direction on the device • Alignment of the sensor housing during initial startup (default setting) as displayed on the ad- jacent figures G00657-01 4.3.2...
  • Page 42: Vertical Flow Direction

    4 Mounting 4.3.4 Vertical flow direction • Vertical installation for measuring abrasive substances, the preferred flow direction is from bottom to top. G00039-01 4.3.5 Horizontal flow direction • Measuring pipe must always be full. • Ensure the line is slightly inclined for degas- sing.
  • Page 43: Installation Near Pumps

    4 Mounting 4.3.8 Installation near pumps • We recommend using mechanical vibration dampers for sensors that are installed near pumps or other equipment generating vibra- tions. G00561 4.3.9 Minimum clearance • Maintain a minimum clearance of 0.7 m (2.3 ft) between the devices to suppress any mutually interference.
  • Page 44: Installation In Pipelines With Larger Nominal Widths

    4 Mounting 4.3.11 Installation in pipelines with larger nominal widths Determine the resulting pressure loss when using reducers (1): 1. Determine the diameter ratio d/D 2. Refer to the flow nomograph for the flow veloc- ity (adjacent figure). 3. Read the pressure loss on the Y-axis in the (1) Flange transition piece flow nomogram.
  • Page 45: 3A-Compliant Mounting

    4 Mounting 4.3.12 3A-compliant mounting • If concentric reducers are mounted on the de- vice, the device must be installed vertically. • Do not mount the device with the terminal box or the transmitter case facing down vertically. • The "mounting bracket" option (1) is not neces- sary here.
  • Page 46: Notes On Opening And Closing The Housing

    4 Mounting Notes on opening and closing the housing To install, start up and maintain the JUMO flowTRANS MAG devices, it is necessary to open the housing of the sensor or transmitter. Be sure to observe the following safety instructions before opening and after closing the housing for re- starting: DANGER –...
  • Page 47: Rotation Of The Transmitter Housing

    4 Mounting 4.4.1 Rotation of the transmitter housing Dual-compartment housing Single-compartment housing Transmitter housing Screw Allen screw 1.  Observe chapter 4.4 "Notes on opening and closing the housing", page 46! 2. Loosen the Allen screws (3) on the front and rear, but do not remove them completely. 3.
  • Page 48: Grounding

    Grounding IMPORTANT (NOTE)! There is the additional document "JUMO flowTRANS MAG S/H - Safety Manual Ex for devices 406012 - 406019" for the measuring systems that are used in potentially explosive areas. The information and data contained in the document must also be strictly observed.
  • Page 49: Metal Pipe With Loose Flanges

    4 Mounting 4.5.3 Metal pipe with loose flanges Using the example of the dual-compartment transmitter housing Flange version Connection flange design Model 406012, 406013, 406015 and 406016 Model 406015 and 406016 Threaded bolts Ground terminal 1. Weld the threaded bolts M6 (1) to the pipeline and establish the ground connection as shown in the illustration.
  • Page 50: Plastic Tubes, Non-Metallic Tubes Or Tubes With Insulating Lining

    4 Mounting 4.5.5 Plastic tubes, non-metallic tubes or tubes with insulating lining Using the example of the dual-compartment transmitter housing IMPORTANT (NOTE)! For plastic lines or pipelines with insulating lining, the measurement medium is to be grounded via the grounding electrodes that must be installed in the device (option). If the pipe section is not free from external interference voltages, it is recommended that one grounding washer is fitted in front of the sensor and one is fitted behind the sensor.
  • Page 51: Installation And Grounding In Pipelines With Cathodic Corrosion Protection (Ccp)

    4 Mounting 4.5.9 Installation and grounding in pipelines with cathodic corrosion protection (CCP) The installation of electromagnetic flowmeters (EMF) in systems with cathodic corrosion protection must be carried out according to the respective system conditions. In particular, the following factors are cru- cial here: a) Pipelines with an electrically conducting or insulating liner b) Pipelines completely or largely on CCP potential or mixed systems with areas on CCP potential and...
  • Page 52 4 Mounting Mixed system pipeline with CCP and functional ground potentials This mixed system has an insulated pipeline with CCP potential and a non-insulated metal pipeline (L = 1/4 x sensor nominal width) with functional ground potential upstream and downstream of the sensor. The following figure shows the preferred installation for cathodic corrosion protection systems.
  • Page 53: Electrical Connection

    Electrical connection 5 Electrical connection Installing the signal and magnet coil cable Water trap Observe the following points when installing cables: • A magnet coil cable (red and brown) is run parallel to the signal lines (violet and blue) so that only one cable is required between the sensor and the transmitter.
  • Page 54: Assembling The Signal And Magnet Coil Cable - Dual-Compartment Housing

    5 Electrical connection Assembling the signal and magnet coil cable - dual-compartment hous- Assemble both cable ends as displayed. IMPORTANT (NOTE)! Use ferrules. • Ferrules 0.75 mm (AWG 19), for the shielding (1S, 2S) • Ferrules 0.5 mm (AWG 20) for all other specifications CAUTION –...
  • Page 55: Assembling The Signal And Magnet Coil Cable - Single-Compartment Housing

    5 Electrical connection Assembling the signal and magnet coil cable - single-compartment housing Assemble the cable end on the transmitter side as displayed in the figure. IMPORTANT (NOTE)! Use ferrules. • Ferrules 0.75 mm2 (AWG 19), for the shielding (1S, 2S) •...
  • Page 56: Sensor Connection

    5 Electrical connection Sensor connection 5.4.1 Metal terminal boxes The connection must only be established when the voltage supply is switched off. The device must be grounded according to the regulations. The sensor must be connected to the trans- mitter via the signal cable (part no. 00648906). IMPORTANT (NOTE)! Use ferrules.
  • Page 57: Protection Type Ip68

    5 Electrical connection 5.4.2 Protection type IP68 WARNING – Impairment of the IP68 protection type! The IP68 protection type of the sensor is impaired by damage to the signal cable. The sheath of the signal cable must not be damaged. Otherwise, the IP68 protection type for the sensor cannot be ensured.
  • Page 58: Connection Via Cable Conduits

    5 Electrical connection 5.4.3 Connection via cable conduits CAUTION – Condensate formation in terminal box! If the sensor is permanently connected to cable conduits, moisture may enter in the terminal box due to condensate formation in the cable conduit. Make sure that the cable inlets on the terminal box are sealed. G01312 A mounting set for sealing the cable conduit is available under part no.
  • Page 59: Potting The Terminal Box

    5 Electrical connection 5.4.4 Potting the terminal box For sensors without Ex-protection or Ex-protection zone, the terminal box can be retroactively potted. A two-component potting compound, which can be ordered separately, is available for retroactive potting of the terminal box on-site. Potting is only possible for sensors mounted horizontally. Observe the following instructions when processing! Preparation Danger –...
  • Page 60: Transmitter Connection

    Transmitter connection IMPORTANT (NOTE)! There is the additional document "JUMO flowTRANS MAG S/H - Safety Manual Ex for devices 406012 - 406019" for the measuring systems that are used in potentially explosive areas. The information and data contained in the document must also be strictly observed.
  • Page 61: Transmitter - Dual-Compartment Housing

    5 Electrical connection 5.5.2 Transmitter - dual-compartment housing Terminal cover The connection terminals for the voltage supply are located under the terminal cover (1). 5.5.3 Transmitter - single-compartment housing G01320 Connection terminals (voltage supply)
  • Page 62: Signal And Magnet Coil Cable Connection

    5 Electrical connection 5.5.4 Signal and magnet coil cable connection Single-compartment transmitter housing For the single-compartment transmitter housing, the outer shielding of the signal and magnet coil cable is connected to the corresponding connection terminal for the signal and magnet coil cable. The shielding for the signal wires functions as a "driven shield"...
  • Page 63: Connection Diagrams

    HART, PROFIBUS-PA and FOUNDATION Fieldbus protocol IMPORTANT (NOTE)! There is the additional document "JUMO flowTRANS MAG S/H - Safety Manual Ex for devices 406012 - 406019" for the measuring systems that are used in potentially explosive areas. The information and data contained in the document must also be strictly observed.
  • Page 64: Electrical Data

    5 Electrical connection Terminal Function 31/32 Current/HART output - The current output is available in "active" or "passive" mode. 97/98 PROFIBUS-PA (PA+/PA-) or FOUNDATION Fieldbus (FF+/FF-) - according to IEC 61158-2 51/52 Digital output DO1 active/passive - Function can be configured using software on-site as "Pulse out- put"...
  • Page 65: Digital Output Do1

    5 Electrical connection 5.7.2 Digital output DO1 1550 1450 1350 1250 1150 1050 ≤ 30 V+ 19 to 21 V+ = 220 mA G00476-04 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 U [V] G00593 Max.
  • Page 66: Digital Input Di1

    5 Electrical connection 5.7.4 Digital input DI1 Ri = 2 k 24V+ G00477-01 Data of the optocoupler 16 V ≤ U ≤ 30 V, R = 2 kΩ e.g. for flowrate to zero or totalizer reset I = internal, E = external 5.7.5 PROFIBUS-PA (PA+/PA-) / FOUNDATION Fieldbus (FF+/FF-) - according to IEC 61158-2...
  • Page 67: Digital Outputs Do1 And Do2

    5 Electrical connection 5.8.2 Digital outputs DO1 and DO2 Separate forward and reverse pulses Separate forward and reverse pulses (connection type) 24V+ G00791 I = internal, E = external 5.8.3 Digital communication PROFIBUS-PA Connection via M12 plug (only in non-potentially explosive areas!) G01003-01 Pin assignment (view from front showing pin insert and pins): PIN 1 = PA+...
  • Page 68 5 Electrical connection...
  • Page 69: Digital Communication

    The communication (configuration, parameterization) can be carried out in conjunction with the DTM (Device Type Manager) available for the device and the corresponding framework applications accord- ing to FDT 1.21. IMPORTANT (NOTE)! The necessary device DTM is included on the CD and can be downloaded from www.jumo.de.
  • Page 70: Profibus-Pa Protocol

    The voltage on the bus line must be in the range between DC 9 and 32 V. For further information  see the separate interface description. 6.2.1 System integration JUMO provides GSD files for integrating devices into the system. IMPORTANT (NOTE)! The necessary DTM device is included on the CD and can be downloaded from www.jumo.de.
  • Page 71: Foundation Fieldbus (Ff)

    DD (device description) file, containing the device description. • CFF (Common File Format) file; this is required for engineering the segment. Engineering can occur online or offline IMPORTANT (NOTE)! The files required for operation are included on the CD, or can be downloaded from www.jumo.de.
  • Page 72 6 Digital communication...
  • Page 73: Startup

    7 Startup IMPORTANT (NOTE)! There is the additional document "JUMO flowTRANS MAG S/H - Safety Manual Ex for devices 406012 - 406019" for the measuring systems that are used in potentially explosive areas. The information and data contained in the document must also be strictly observed.
  • Page 74: Transmitter - Dual-Compartment Housing

    7 Startup 7.2.1 Transmitter - dual-compartment housing JUMO flowTRANS MAG S/H - compact design JUMO flowTRANS MAG - remote mount design Case lid Transmitter plug-in module Mounting screws Backplane (in the transmitter housing) Jumper (BR901) for active/passive current output Jumper (BR903) for integral/remote mount design...
  • Page 75: Transmitter - Single-Compartment Housing

    7 Startup 7.2.2 Transmitter - single-compartment housing JUMO flowTRANS MAG S/H - compact design JUMO flowTRANS MAG - remote mount design Case lid Transmitter plug-in module Mounting screws Backplane (in the transmitter housing) Jumper (BR905, BR906) for communication Jumper (BR901) for active/passive current output...
  • Page 76: Startup Of Profibus-Pa Devices

    7 Startup Jumper Position Function BR901 active Current output 31/32 active passive Current output 31/32 passive BR902 Read only Hardware write protection active BR903 Integral Transmitter with compact design Remote Transmitter with remote mount design BR904 active Pulse output 52/52 active passive Pulse output 52/52 passive BR905, BR906...
  • Page 77: Setting The Address Locally For Transmitters With A Dual-Compartment Housing

    7 Startup 7.3.1 Setting the address locally for transmitters with a dual-compartment housing Transmitter plug-in module DIP switch Switch assignment Switch Assignment 1 to 7 PROFIBUS address Defining the address mode: Off = The address is set via the bus (default setting) On = The address is set via the DIP switches 1 to 7 (locally) Device behavior on switching on the voltage supply After the voltage supply is switched on, DIP switch 8 is queried:...
  • Page 78: Configuration For Transmitters With A Single-Compartment Housing

    7 Startup 7.3.2 Configuration for transmitters with a single-compartment housing JUMO flowTRANS MAG S/H - compact design JUMO flowTRANS MAG - remote mount design Case lid Transmitter plug-in module Mounting screws Backplane (in the transmitter housing) Jumper (BR905, BR906) for communication...
  • Page 79: Voltage/Current Consumption

    1 x DO and all manufacturer-specific parameters The manufacturer-specific GSD file "JUMO_3430.gsd" is available for download on the manufacturer homepage at www.jumo.de. The standard GSD files "PA1397xx.gsd" are available for download on the homepage of PROFIBUS In- ternational at http://www.profibus.com.
  • Page 80: Startup For Foundation Fieldbus Devices

    The CFF file is required for engineering the segment. Engineering can occur online or offline. IMPORTANT (NOTE)! The files required for operation are included on the CD, or can be downloaded from www.jumo.de. The device's FOUNDATION Fieldbus interface complies with the standards FF-890/891 and FF-902/90.
  • Page 81: Configuration For Transmitters With A Dual-Compartment Housing

    7 Startup 7.4.1 Configuration for transmitters with a dual-compartment housing Transmitter plug-in module DIP switch Switch assignment Switch Assignment Release of AI function block simulation Hardware write protection for write access via the bus (all blocks locked) Switch Status Simulation Mode Write Protect Disabled Disabled...
  • Page 82: Configuration For Transmitters With A Single-Compartment Housing

    7 Startup 7.4.2 Configuration for transmitters with a single-compartment housing JUMO flowTRANS MAG S/H - compact design JUMO flowTRANS MAG - remote mount design Case lid Transmitter plug-in module Mounting screws Backplane (in the transmitter housing) Jumper (BR905, BR906) for communication...
  • Page 83: Setting The Bus Address

    7 Startup 7.4.3 Setting the bus address The bus address for the FOUNDATION Fieldbus is assigned automatically by the LAS (Link Active Scheduler). Address detection relies on a unique number (DEVICE_ID). The number is composed of the manufacturer ID, device ID and device serial number. The switch-on behavior is compliant with draft DIN IEC/65C/155/CDV from June 1996.
  • Page 84: Starting Up The Device

    7 Startup Starting up the device 7.5.1 Loading the system data 1. Switch on the voltage supply. After switching on the voltage supply, the following messages are dis- played consecutively in the LCD display: 2. Download the system data as follows: For a completely new system or initial startup The calibration data of the sensor and the transmitter settings are loaded from the SensorMemory into the transmitter.
  • Page 85: Error Message "Incompatible Sensor

    The display for the current flow is ZERO. The current output assumes its pre-con- figured state (Iout for alarm). Make sure that the sensor and the transmitter are of the same device version (e.g., sensor 406012/2-0 JUMO flowTRANS MAG S, transmitter 406018/2-0 JUMO flowTRANS MAG S).
  • Page 86: Parameterization Via The "Easy Setup" Menu Function

    7 Startup 7.5.3 Parameterization via the "Easy Setup" menu function Upon request, the device can be parameterized at the factory according to customer specifications. If no customer information is available, the device is delivered with default settings. The setting of the most common parameters is summarized in the "Easy Setup" menu. This menu pro- vides the fastest way to configure the device.
  • Page 87 7 Startup Easy Setup Language German Next Edit 7. Use to call up the edit mode. 8. Use to select the desired language. 9. Use to confirm your selection. Easy Setup Q (Flowrate) Unit Next Edit 10. Use to call up the edit mode. 11.
  • Page 88 7 Startup 19. Use to call up the edit mode. 20. Use to select the desired operating mode. "Pulse Mode": in pulse mode, pulses per unit are output. The settings for this are provided in the next menu. "Fullscale Frequency": in frequency mode, a frequency proportional to the flow is output. The maxi- mum frequency can be configured according to the flow measuring range.
  • Page 89 7 Startup 31. Use to call up the edit mode. 32. Use to select the desired alarm mode. 33. Use to confirm your selection. Easy Setup Low Alarm Value 3.5000 mA Next Edit 34. Use to call up the edit mode. 35.
  • Page 90 7 Startup Entry of the signal cable length between the transmitter and the sensor. For compact devices, 0.01 m must be entered. 43. Use to confirm your setting. Menu Easy Setup Exit Select 44. Use to switch to the process display The main menu is displayed again after all parameters are set.
  • Page 91 Low Flow Cut Off 0 to 10 % Empty Pipe Detector ON/OFF TFE Detector ON/OFF Only available for JUMO flowTRANS MAG S02/H02 with the PROFIBUS-PA version Possible parameterization Default setting PA Addr (-BUS-) 0 to 126 Ident Nr. Selector...
  • Page 92: Nominal Width, Measuring Range

    7 Startup Nominal width, measuring range The measuring range end value can be configured between 0.02xQ DN and 2xQ Nominal width Minimum Maximum Measuring range end value 0 to ≈ 10 m/s Measuring range end value 0.02xQ DN (≈ 0.2 m/s) DN (≈...
  • Page 93 7 Startup Nominal width Minimum Maximum Measuring range end value 0 to ≈ 10 m/s Measuring range end value 0.02xQ DN (≈ 0.2 m/s) DN (≈ 20 m/s) DN (inch) US gal/min US gal/min US gal/min 1400 (54) 1080 4755 54000 237755 108000...
  • Page 94 7 Startup...
  • Page 95: Parameterization

    Parameterization 8 Parameterization Operation The LCD display is provided with capacitive push-buttons. They allow you to operate the device through the closed lid. IMPORTANT (NOTE)! The transmitter automatically calibrates the capacitive push-buttons on a regular basis. If the housing is opened during operation, the sensitivity of the push-buttons is first increased meaning that operating er- rors may occur.
  • Page 96: Menu Levels

    8 Parameterization Menu levels Process display Information level Configuration level Diagnostics Easy Setup Device Info Operator Page 1 ... 3 Autoscroll Device Setup Signals View Display Input/Output Process Alarm Communication Diagnostics Totalizer Process display The process display shows the current process values. Information level The information level contains the parameters and information that are relevant for the user.
  • Page 97: Process Display

    8 Parameterization 8.2.1 Process display G00756 Measuring point identifier Current process values Symbol indicating the button function Symbol indicating "Parameterization protected" The process display appears on the LCD display when the device is switched on. It displays information about the device and current process values. The display of the current process values (2) can be adjusted at the configuration level.
  • Page 98: Switch To The Information Level (Operator Menu)

    8 Parameterization 8.2.2 Switch to the Information level (Operator Menu) In the Information level, you can use the Operator Menu to display diagnostics information and select which operator pages are displayed. Process display 1. Use to switch to the Information level. Operator Menu Diagnostics Operator Page 1...
  • Page 99 8 Parameterization Error messages in the LCD display In case of an error, a message consisting of a symbol and text (e.g. Electronics) appears at the bottom of the process display. The displayed text provides information about the area in which the error has oc- curred.
  • Page 100 8 Parameterization Calling up the error description Further information about the occurred error can be called up at the Information level. Process display Electronics 1. Use to switch to the Information level. Operator Menu Diagnostics Operator Page 1 Operator Page 2 Back Select 2.
  • Page 101: Switching To The Configuration Level (Parameterization)

    8 Parameterization 8.2.3 Switching to the Configuration level (Parameterization) The device parameters can be displayed and changed at the Configuration level. Process display 1. Use to switch to the Configuration level. Access Level Read Only Standard Advanced Next 2. Use to select the Access Level.
  • Page 102: Hardware Write Protection

    In addition to the password protection, you can also activate hardware write protection. Transmitter in a dual-compartment housing Transmitter in a single-compartment housing JUMO flowTRANS MAG S/H - compact design JUMO flowTRANS MAG - remote mount design Case lid Jumper (BR902) for hardware write protection 1.
  • Page 103: Selecting And Changing Parameters

    8 Parameterization 8.2.5 Selecting and changing parameters Table entry When an entry is made from a table, a value is selected from a list of parameter values. Menu name Parameter name Value currently set Next Edit 1. Selection of the parameters to be set in the menu. 2.
  • Page 104 8 Parameterization Alphanumeric entry When an alphanumeric entry is made, a value is set by entering the individual decimal positions. -----Menu name---- Parameter name Value currently set Next Edit 1. Selection of the parameters to be set in the menu. 2.
  • Page 105: Parameter Overview In The Configuration Level

    8 Parameterization Parameter overview in the configuration level IMPORTANT (NOTE)! This parameter overview shows all the menus and parameters available on the device. Not all of the menus and parameters may be visible on the device, depending on the device equipment and configuration.
  • Page 106 8 Parameterization Analog Range Amplifier Adjust CMReject CMR Value Adjust Gain 1...64 Adjust Gain 1 Adjust Gain 8 Adjust Gain 32 Adjust Gain 64 Transmitter Device Version ScanMaster Option Tx Typ Tx Spanne Tx Nullpunkt Offset Iout Gain Iout Simulator Transmitter ID SAP / ERP Nr.
  • Page 107 8 Parameterization Operating Mode Meter Mode Flow Indication System Zero Manual Adjust Auto Adjust Noise Reduction Mean Filter Notch Filter Lowpass V=Auto Lowpass V=1 Display Language Contrast Operator Pages Operator Page 1 Display Mode 1st Line Autoscroll Operator Page 2 2nd Line Flowrate Format 3st Line...
  • Page 108 8 Parameterization...
  • Page 109 8 Parameterization...
  • Page 110 8 Parameterization...
  • Page 111 8 Parameterization...
  • Page 112 8 Parameterization...
  • Page 113: Parameter Description

    8 Parameterization Parameter description 8.4.1 Menu: Easy Setup Menu/parameter Value range Description Easy Setup Menu "Easy Setup" Language Deutsch, English, Français, Menu language selection Español, Italiano, Dansk, Svenska, Polski, Russki, Zhongweng, Turkce Q (Flowrate) Unit l/s; l/min; l/h; ml/s; ml/min; Unit selection for the flow display m3/s;...
  • Page 114 Italics = Parameter can only be viewed in the "Advanced" password level Cable length The input is required here for devices of the JUMO flowTRANS MAG S02/H02 design type to be able to use the diagnostic function. If the ScanMaster verification software is used, the signal cable length must also be input.
  • Page 115: Menu: Device Info

    8 Parameterization 8.4.2 Menu: Device info IMPORTANT (NOTE)! This menu is only used for displaying the device parameters. The parameters can be viewed inde- pendently of the set access level, however cannot be changed. Menu/parameter Value range Description Device Info Sensor Selection of "Sensor"...
  • Page 116 8 Parameterization Properties Selection of "Properties" submenu Italics = Parameter can only be viewed in the "Advanced" password level You toggle between the measuring ranges using the digital input or via the "Config. Device/Sensor/Dual ranges" menu Menu/parameter Value range Description Device Info/Sensor/Calibration.
  • Page 117 8 Parameterization Menu/parameter Value range Description Device Info/Acquisition Rate ADC Display for service purposes only Analog Reset Selection of "Analog Reset" submenu Driver Selection of "Driver" submenu Italics = Parameter can only be viewed in the "Advanced" password level Menu/parameter Value range Description Device Info/Acquisition/Analog Reset...
  • Page 118 8 Parameterization Menu/parameter Value range Description Device Info/Transmitter Device Version Display of the transmitter device version ScanMaster option Display indicating whether the ScanMaster option is enabled. TX Type Display of the transmitter type TX Span Calibration value of the transmitter (range) TX Zero Calibration value of the transmitter (zero point) Offset lout...
  • Page 119: Menu: Device Setup

    8 Parameterization 8.4.3 Menu: Device Setup Menu/parameter Value range Description Device Setup ... Access Control Selection of "... Access Control" submenu ... Sensor Selection of "... Sensor" submenu ... Transmitter Selection of "... Transmitter" submenu Menu/parameter Value range Description Device Setup/... Access Control Standard Password Alphanumeric Enter/change the password for the "Standard"...
  • Page 120 8 Parameterization Menu/parameter Value range Description Device Setup/... Transmitter ... Unit Selection of "Unit" submenu Damping 0.02 to 60 s Damping setting The value set here refers to 1 t (tau). The specification refers to the response time for a sudden change in the flow.
  • Page 121 8 Parameterization Menu/parameter Value range Description Device Setup/... Transmitter/... Unit Totalizer/Pulse Unit m3, l, ml, ft3, hl, g, kg, t, lb, ig- Unit selection for the flow counter al, ugal, bls, Ml, Mugal, Custom/s Q (Flowrate) Unit l/s, l/min, l/h, ml/s, ml/min, Unit selection for the flow display m3/s, m3/min, m3/h, m3/d, Ml/d, ft3/s, ft3/min, ft3/h, ft3/d,...
  • Page 122 8 Parameterization Menu/parameter Value range Description Device Setup/... Transmitter/... Low Flow Cut Off Threshold 0 to 10 % Select the switching threshold for monitoring the low flow If the value falls below the set switching threshold, no flow is measured. The current output is set to zero. The switching threshold of the low flow monitoring re- fers to the measuring range currently set.
  • Page 123: Menu: Display

    8 Parameterization 8.4.4 Menu: Display Menu/parameter Value range Description Display Language Deutsch, English, Français, Menu language selection Español, Italiano, Dansk, Svenska, Polski, Russki, Zhongweng, Turkce Contrast 0 to 100 % Contrast setting for the LCD display Operator Pages Selection of "Operator Pages" submenu Autoscroll ON/OFF When Multiplex mode is enabled, you can activate...
  • Page 124 Current Output [mA] SignalProportion Reference Signal Max. Signal Min. Amplification Noise Reset Count. Total Batch Count. Batch Totalizer Conductivity Sensor Temp Only available for JUMO flowTRANS MAG S02/H02 Italics = Parameter can only be viewed in the "Advanced" password level...
  • Page 125: Menu: Input/Output

    Default setting: Flowrate to Zero Current Output Selection of "Current output" submenu Only available for JUMO flowTRANS MAG S02/H02 Italics = Parameter can only be viewed in the "Advanced" password level Digital Input setup If the filling action is stopped before reaching the configured filling quantity, the filling counter is reset to...
  • Page 126 8 Parameterization Menu/parameter Value range Description Input/Output/... Digital Out Mode Function DO1/DO2 Pulse F/Pulse R Selection of functions for the digital outputs DO1 and Pulse F/Logic Pulse F/Pulse R: Pulse FR/Logic Logic/Logic • DO1 = Pulse output for the forward direction •...
  • Page 127 "DO2 Signal" parameter. DO2 Action Normally Open, Normally Selection of the switching behavior for the digital out- Closed Default setting: Normally Open Only available for JUMO flowTRANS MAG S02/H02 Italics = Parameter can only be viewed in the "Advanced" password level...
  • Page 128 Conductivity ON/OFF Coating ON/OFF Sensor Temp ON/OFF Only available for JUMO flowTRANS MAG S02/H02 Menu/parameter Value range Description Input/Output/... Logic Setup/... DO2 Alarm Config See description "..DO1 Alarm Config" Italics = Parameter can only be viewed in the "Advanced" password level...
  • Page 129 8 Parameterization Menu/parameter Value range Description Input/Output/Current Output lout at Alarm High Alarm, Low Alarm Selection of the status for the current output in case of a fault The output "low" or "high" current is set in the follow- ing menu. Default setting: High Low Alarm Value 3.5 to 3.6 mA...
  • Page 130: Menu: Process Alarm

    8 Parameterization 8.4.6 Menu: Process alarm Menu/parameter Value range Description Process Alarm Clear Alarm History The alarm list can be cleared here. Group Masking Selection of "Group Masking" submenu Individual Masking Selection of "Individual Masking" submenu Alarm Simulation OFF ... Different alarm messages and output statues can be simulated.
  • Page 131: Menu: Communication

    8 Parameterization 8.4.7 Menu: Communication Menu/parameter Value range Description Communication HART Selection of "HART" submenu Cyclic Data Out Select the "Cyclic. Data Out" submenu Service Port Selection of "Service Port" submenu PROFIBUS Selection of "PROFIBUS" submenu The menu is only displayed for devices with PROFI- BUS-PA.
  • Page 132 8 Parameterization Menu/parameter Value range Description Communication/Cyclic Data Out/Cyclic Upd. Select Flow Group ON/OFF Selection of the data to be output via the infrared ser- vice port Contents: Q (%), Q (l/s), v (m/ The diagnostics data is consolidated into groups. Each group can be separately switched on or off and Output Group ON/OFF...
  • Page 133 The channel can be selected with the "Al4 Channel" parameter. To display active values, switch on sensor measure- ments, and/or conductivity measurements. Al4-Channel Sensor Temp, Conductivity Selection of the channel output by AI4 The PV_SCALE and OUT_SCALE structure is not adjusted. Only available for JUMO flowTRANS MAG S02/H02...
  • Page 134 Zero Adjust, Totalizer Stop (all), Dual range , Start/Stop Batching Only available for JUMO flowTRANS MAG S02/H02 Italics = Parameter can only be viewed in the "Advanced" password level PA Addr (-BUS-) DIP switch (only for transmitter in a dual-compartment housing): •...
  • Page 135 8 Parameterization Menu/parameter Value range Description Communication/FF The menu is only displayed for devices with FOUN- DATION Fieldbus. Show FF Address Display only Display the FOUNDATION Fieldbus address. The address is set via the FOUNDATION Fieldbus mas- ter. AI-Q Flowrate Display only Current flow in the set unit from the transducer block flow including status.
  • Page 136: Menu: Diagnostics

    Grounding Check Selection of "Grounding check" submenu TFE Detector Selection of "TFE Detector" submenu SIL Detection Selection of "SIL Detection" submenu Only available for JUMO flowTRANS MAG S02/H02 Italics = Parameter can only be viewed in the "Advanced" password level...
  • Page 137 8 Parameterization Menu/parameter Value range Description Diagnostics/Diagnostics Control/Empty Pipe Detector Empty Pipe Detector ON/OFF Preconditions for activating the "Empty Pipe Detec- tor" function: • Only for nominal widths ≥ DN 10 and without pre- amplifier • There must be no flow through the sensor (close all valves, shut-off devices, etc.).
  • Page 138 Setting of the minimum limit value for the sensor tem- perature. When this value is not reached, an alarm is triggered. Default setting: -50 C° Only available for JUMO flowTRANS MAG S02/H02 Italics = Parameter can only be viewed in the "Advanced" password level...
  • Page 139 Gas Bubble Value Display only Display the current Gas Bubble Value Only available for JUMO flowTRANS MAG S02/H02 Italics = Parameter can only be viewed in the "Advanced" password level Gas Bubble Detector The gas bubble detector can be used in the nominal width range DN 10 to 300. More information see chapter 9 "Advanced diagnostics functions", page 153.
  • Page 140 Sets the maximum limit value for impedance. When the value exceeds this limit, an alarm is triggered. Default setting: 20,000 Ω Only available for JUMO flowTRANS MAG S02/H02 Italics = Parameter can only be viewed in the "Advanced" password level Cond. Detection The conductivity measurement can be used in the nominal width range DN 10 to 300.
  • Page 141 8 Parameterization Menu/parameter Value range Description Diagnostics/Diagnostics Control/TFE Detector TFE Detector Activation of the "Partial filling detector" function (TFE) Adjust TFE Full Partial filling detection must be calibrated to match the local conditions. Starts the automatic calibration of partial filling detec- tion (see note) Cross-reference TFE Threshold Manual fine adjustment of the switching threshold...
  • Page 142 On Demand F.P. Selection of the "On Demand F.P." submenu Only available for JUMO flowTRANS MAG S02/H02 The fingerprint database built into the transmitter can be used to compare the values at the time of fac- tory calibration or startup with the values currently acquired.
  • Page 143 Start Tx Startup F.P. Creates the Startup Fingerprint for the transmitter. Tx Check CMR Display only Tx Check 5 m/s Tx Check 10 m/s Only available for JUMO flowTRANS MAG S02/H02 Menu/parameter Value range Description Diagnostics/Fingerprints/On Demand F.P. Start On Demand FP...
  • Page 144 8 Parameterization Menu/parameter Value range Description Diagnostics/Flowrate Alarm Max. Flowrate Alarm 0 to 130 % Setting of the maximum limit value for the flow. Min. Flowrate Alarm 0 to 130 % Setting of the minimum limit value for the flow. Italics = Parameter can only be viewed in the "Advanced"...
  • Page 145: Menu: Totalizer

    Selection of "... Reset Vol. Totals" submenu ... Batching Selection of "... Batching" submenu ... Edit Vol. Totals Selection of "... Edit Vol. Totals" Only available for JUMO flowTRANS MAG S02/H02 Menu/parameter Value range Description Totalizer/... Reset Vol. Totals Edit Totalizer Fwd...
  • Page 146 Manual start/stop of the filling action. Alternatively, the digital input can be used to start/stop the filling ac- tion. Only available for JUMO flowTRANS MAG S02/H02 Italics = Parameter can only be viewed in the "Advanced" password level Menu/parameter Value range Description Totalizer/Totalizer Default Setting.
  • Page 147: Alarm Simulation

    8 Parameterization Alarm Simulation Different alarms can be simulated in the "Process alarm/alarm simulation" menu. Menu/Parameter Description Process alarm/Alarm simulation Alarm simulation switched off 0-Sim.CurrentOut Simulate current output 1-Sim.Logic on DO1 Switch on/off switching output (terminal 51/52) 2-Sim.Pulse on DO1 Simulate pulse output (terminal 51/52) 3-Sim.Logic on DO2 Switch on/off switching output (terminal 41/42)
  • Page 148 8 Parameterization Menu/Parameter Description 43-Gas bubble Simulate "Gas bubbles" error 44-Pulse Cut Off Simulate "Pulse output" error 46-Sensor Temp Simulate "Sensor temperature alarm" error...
  • Page 149: Filling Operation - Jumo Flowtrans Mag S02/H02

    Digital input Digital output The integrated filling function (batch) on the models JUMO flowTRANS MAG S02/H02 lets you control filling actions with a filling time > 3 seconds. To allow this to happen, the filling quantity is defined using a configurable counter.
  • Page 150: Configuration

    Function DO1/DO2 Logic/Logic Selects the operating mode for the Digital Input Setup • Start/Stop Batching (filling function) Only available for JUMO flowTRANS MAG S02/H02 The following settings must be configured to actuate the filling valve: Menu/parameter Value range Description Input/output...
  • Page 151 Alternatively, the digital input can be used to start/ stop the filling action. Only available for JUMO flowTRANS MAG S02/H02 To display the data relevant to filling in the process display, one of the operator pages should be config- ured accordingly in the "Display" menu:...
  • Page 152 8 Parameterization...
  • Page 153: Advanced Diagnostics Functions

    For this function, the conductivity of the measurement medium must be between 20 µS/ cm and 20,000 µS/cm. • The function is only available for JUMO flowTRANS S/H devices without Ex-protection or with Ex- protection for zone 2. Additional installation conditions: •...
  • Page 154: Electrode Coating Detection

    When the configured threshold is exceeded, an alarm is triggered via the programmable output, depend- ing on the configuration. Conditions for using the function: • JUMO flowTRANS MAG S02: The function is available in the nominal width range DN 10 to 300 (3/8 to 12 inch). • JUMO flowTRANS MAG H02: The function is available in the nominal width range DN 10 to 100 (3/8 to 4 inch).
  • Page 155: Electrode Impedance Monitoring

    If the value exceeds or drops below the configured threshold, an alarm is triggered via the programmable output, depending on the configuration. Conditions for using the function: • JUMO flowTRANS MAG S02: The function is available in the nominal width range DN 10 to 300 (3/8 to 12 inch). • JUMO flowTRANS MAG H02: The function is available in the nominal width range DN 10 to 100 (3/8 to 4 inch).
  • Page 156: Fingerprint

    Display only Amplitude 3 Display only Amplitude 4 Display only Only available for JUMO flowTRANS MAG S02/H02 1. Use to select the "Grounding Check" item. 2. Use to start the "Grounding Check" function. After starting the grounding check, the frequency range up to 250 Hz is measured. The 4 strongest...
  • Page 157 9 Advanced diagnostics functions The matching amplitudes and the power spectrum across the frequency range can be displayed with the following parameters. 3. Use to select the "Power Spectrum" item. 4. Use to display the parameter. 5. Use to select the "Amplitude 1 Value (n)" item. 6.
  • Page 158: Recommended Settings For Diagnostics Thresholds

    9 Advanced diagnostics functions Recommended settings for diagnostics thresholds Thresholds for the diagnostics measured values can be entered in the "Diagnostics/Diagnostics Con- trol/..." menu. To facilitate the settings, recommended settings for the individual thresholds are listed here. The stated values are only to be understood as rough guidelines which may need to be adapted to match the local conditions.
  • Page 159: Electrode Coating Thresholds

    9 Advanced diagnostics functions 9.3.2 Electrode coating thresholds Electrode deposit monitoring is disabled as a default setting and can be enabled in the "Diagnostics/ Diagnostics Control/Coating Detector" menu. Parameter Default setting Coating QE min. alarm 0.0 Ohm Coating QE max. alarm 100,000 Ohm Recommended setting in menu "Diagnostics/Diagnostics Control/Coationg Detector": •...
  • Page 160 9 Advanced diagnostics functions...
  • Page 161: 10 Error Messages

    Error messages 10 Error messages 10.1 Calling up the error description Further information about the occurred error can be called up at the Information level. Process display Electronic unit 1. Use to switch to the Information level. Operator menu Diagnostics ..Operator Page 1 ..Operator Page 2 Back...
  • Page 162: Function Check

    10 Error messages Error no./ Text in the LCD display Cause Remedy area F251.040 Self Check Alarm The SIL monitor function detected Replace transmitter or contact Electronics an error in the transmitter. manufacturer F250.016 Tx memory fault detected Error in the transmitter electronics Replace electronics or contact Electronics unit...
  • Page 163 10 Error messages Error no./ Text in the LCD display Cause Remedy area C190.045 An alarm is simulated Simulation pulse output is In the "Diagnostics" menu, acti- Config. Switch off alarm simulation switched on. vate the simulation mode. C186.009 Tx Simulator/Calibrator Transmitter is operated on the In the "Diagnostics"...
  • Page 164 10 Error messages Error no./ Text in the LCD display Cause Remedy area C164.003 Logic Simulation selected The simulation mode is switched In the "Process Alarm" menu, Config. on DO2 switch off the simulation mode. Switch off Simulation Mode C158.039 Simulation of HART fre- The simulation mode is switched In the "Process Alarm"...
  • Page 165: Operating The Device Out Of Spec

    10 Error messages 10.2.3 Operating the device out of spec Error no./ Text in the LCD display Cause Remedy area S149.021 Coil resistor out of limits Coil resistance too high: coil or coil Check wiring, check coil circuit Operation circuit fuse is defective or M1/M2 fuse, contact manufacturer Check wiring wired incorrectly or cable break or...
  • Page 166 10 Error messages Error no./ Text in the LCD display Cause Remedy area S110.035 Sensor setup Cal-Status Sensor is uncalibrated or Cal-Sta- Contact manufacturer Operation tus is not set to "calibrated". Set Cal-Status to calibrated S108.044 Pulse output is cutted off Incorrect configuration In the "Easy Setup"...
  • Page 167: Maintenance

    10 Error messages 10.2.4 Maintenance Error no./ Text in the LCD display Cause Remedy area M099.027 NV Corrupt NV Memory, SensorMemory, Contact manufacturer Electronics FRAM defective M094.034 Current out fault 20 mA loop open, cable break or Check for incorrect wiring or wire Electronics no power is connected during op- break.
  • Page 168 10 Error messages Error no./ Text in the Current Digital Pulse Display Error area LCD display output output output can be behavior behavior behavior masked? F244.031 Internal supply voltage er- lout at alarm General Alarm 0 Hz Electronics Contact service F236.024 DC to High lout at alarm...
  • Page 169 10 Error messages Error no./ Text in the Current Digital Pulse Display Error area LCD display output output output can be behavior behavior behavior masked? C172.04 Pulse Simulation selected Current value No response Simulated Current Mask group Configuration on DO2 value value Switch off Simulation Mode...
  • Page 170 10 Error messages Error no./ Text in the Current Digital Pulse Display Error area LCD display output output output can be behavior behavior behavior masked? S120.023 Electrode Noise too high Current value No response Current Current Mask group Operation value value Switch on Noise reduction S110.035...
  • Page 171: Error Message During Startup

    10 Error messages 10.3.1 Error message during startup No sensor detected -System Startup- NO SENSOR DETECTED Device Resetting Offline Once the device is switched on, the calibration data of the sensor and the transmitter settings is loaded from the SensorMemory into the transmitter. If the communication cannot be established to the Sen- sorMemory , the adjacent message appears on the LCD display.
  • Page 172 The display for the current flow is ZERO. The current output assumes its pre-con- figured state (Iout for alarm). Make sure that the sensor and the transmitter are of the same device version (e.g., sensor 406012/2-0 JUMO flowTRANS MAG S, transmitter 406018/2-0 JUMO flowTRANS MAG S).
  • Page 173: 11 Maintenance

    All repair or maintenance work must only be carried out by qualified customer service personnel. When replacing or repairing individual components, original spare parts must be used. To install, start up and maintain the JUMO flowTRANS MAG devices, it is necessary to open the housing of the sensor or transmitter.
  • Page 174: Sensor

    11 Maintenance 11.2 Sensor The sensor is largely maintenance-free. The following items must be checked annually: • Environmental conditions (ventilation, moisture) • Leak-tightness of process connections • Cable inlets and lid screws • Functional safety of the voltage supply, lightning protection and station ground The sensor electrodes must be cleaned if the flow information on the transmitter changes when record- ing the identical flow volume.
  • Page 175: Replacing The Transmitter Or Sensor

    IMPORTANT (NOTE)! When replacing the transmitter or sensor, the correct assignment must be ensured. The respective series is specified on the nameplate of the transmitter or sensor (e.g. JUMO flowTRANS MAG S or JUMO flowTRANS MAG H). After the transmitter is replaced, the system data must be reloaded according to the specifications in the operating manual (...
  • Page 176: Replacing The Transmitter

    11 Maintenance 11.4.2 Replacing the transmitter Transmitter in a dual-compartment housing Transmitter in a single-compartment housing Case lid Transmitter plug-in module Mounting screws Lid lock Replace the transmitter plug-in as follows: 1.  Observe chapter 11.1 "General information", page 173! 2.
  • Page 177: 12 Spare Parts List

    Spare parts list 12 Spare parts list 12.1 Transmitter electronics unit fuse Designation Part number Fuse (1.0 A) for voltage supply; usable for all devices 00645917 12.2 Spare parts for devices with a compact design Designation Part number Housing lid for transmitters in a single-compartment housing with a com- 00645918 pact design Universal backplane for transmitters in a dual-compartment housing...
  • Page 178: Spare Parts For Devices With A Remote Mount Design

    12 Spare parts list 12.3 Spare parts for devices with a remote mount design 12.3.1 Field housing Designation Part number Front housing lid for transmitters in a dual-compartment housing with a re- 00645925 mote mount design (ex-zone 1) Universal backplane for transmitters in a dual-compartment housing 00645920 Housing lid for transmitters in a single-compartment housing with a re- 00645928...
  • Page 179: Sensor - Zone 2

    12 Spare parts list 12.3.2 Sensor - zone 2 G00880 Designation Part number Model 406012 Model 406015 Model 406013 Model 406016 Connection board (without preamplifier) 00645931 00645931 Connection board (with preamplifier) 00645932 00645932 12.3.3 Sensor - zone 1 G00877 Designation Part number Model 406012 Model 406013...
  • Page 180: 3A-Compliant Seal Materials

    12 Spare parts list 12.4 3A-compliant seal materials 12.4.1 Flat seals EPDM Designation Part number Flat seal with FDA approval 00661158 Nominal width: DN 3 to 10 (1/10″ to 3/8″) Flat seal with FDA approval 00661173 Nominal width: DN 15 (1/2″) Flat seal with FDA approval 00661174 Nominal width: DN 20 (3/4″)
  • Page 181: 13 Technical Data

    30 min 13.1.2 Maximum measurement deviation IMPORTANT (NOTE)! DN  See table chapter 7.6 "Nominal width, measuring range", page 92. JUMO flowTRANS MAG S01 Calibration (pulse output) Nominal width ±0.4% of the measured value, ±0.02% QmaxDN DN 3 to 2,000 (standard) ±0.2% of measured value, ±0.02% QmaxDN (option-...
  • Page 182: Repeatability, Response Time

    13 Technical data ±0.2% of measured value, ±0.02% QmaxDN (option- DN 10 to 100 Flow velocity v in [m/s], Q/Q Accuracy ± of measured value in [%] DN  See table chapter 7.6 "Nominal width, measuring range", page 92. Analog output effect Same as pulse output plus ±0.1% of measured value + 0.01 mA 13.1.3 Repeatability, response time...
  • Page 183: Sensor Jumo Flowtrans Mag S01/02

    13 Technical data 13.2 Sensor JUMO flowTRANS MAG S01/02 13.2.1 Temperatures The temperature range of the device depends on a series of factors. These factors include the measure- ment medium temperature, the ambient temperature, the operating pressure, the lining material, and the approvals for the Ex-protection.
  • Page 184: Maximum Ambient Temperature Depending On The Medium Temperature

    Maximum ambient temperature depending on the medium temperature IMPORTANT (NOTE)! There is the additional document "JUMO flowTRANS MAG S/H - Safety Manual Ex for devices 406012 - 406019" for the measuring systems that are used in potentially explosive areas. The information and data contained in the document must also be strictly observed.
  • Page 185: Protection Type According To En 60529

    13 Technical data Remote mount design Lining Flange Ambient temperature Medium temperature material Minimum Maximum Minimum Maximum 80 °C (176 °F) Steel -10 °C (14 °F) 60 °C (140 °F) -10 °C (14 °F) 90 °C (194 °F) Hard rubber 80 °C (176 °F) Stainless steel -15 °C (5 °F)
  • Page 186: Insertion Length

    13 Technical data 13.2.8 Insertion length The flanged devices comply with the insertion lengths specified according to VDI/VDE 2641, ISO 13359 or according to DVGW (work sheet W420, design WP, ISO 4064 short). 13.2.9 Signal cable length and preamplifier For devices in the remote mount design, the electrical connection between the transmitter and sensor is made via a signal cable.
  • Page 187: Sensor Materials

    Cable fitting with M20 × 1.5 or NPT thread For Ex-version for -40 °C (40 °F) ambient temperature Measuring pipe materials  see chapter 13.2.12 "Measuring pipe materials", page 189. Only for JUMO flowTRANS MAG S02 version Process connection materials see chapter 13.2.11 "Process connection materials", page 189.
  • Page 188 13 Technical data Steel sensors Component Standard Option touched by me- dium Housing for nominal width: DN 25 to 600 (1 to 24″) Steel, painted, Paint coat, ≥ 80 µm thick, RAL 9002 (gray white) Terminal box Alum. alloy, painted, ≥...
  • Page 189: Process Connection Materials

    13 Technical data 13.2.11 Process connection materials Stainless steel 1.4301 (AISI 304), 1.4307, 1.4404 (AISI 316 L) 1.4435 (AISI 316 L), 1.4541 (AISI 321), 1.4571 (AISI 316 Ti), ASTM A182 F304, ASTM A182 F304 L, ASTM 182 F316 L, ASTM A182 F321, ASTM A182 F316 TI, ASTM A 182 F316, 0Cr18Ni9, 0Cr18Ni10, 0Cr17Ni13Mo2, 0Cr27Ni12Mo3, 1Cr18Ni9Ti, 0Cr18Ni12Mo2Ti Steel (painted), steel (zinc-plated)
  • Page 190 13 Technical data Steel DIN flange Steel DIN flange up to DN 600 (24") DN 700 to 1,000 (28 to 40") [bar] [PSI] 1595.4 1450.4 PN 100 1305.3 1160.3 1015.3 PN 63 870.2 725.2 580.2 PN 40 435.1 PN 25 290.1 PN 16 145.0...
  • Page 191 13 Technical data Steel sensors Limits for the admissible fluid temperature (TS) and admissible pressure (PS) are determined by the used liner and flange materials of the device (see device nameplate). Cast steel case Welded steel case DN 25 to 600 (1 to 24") DN 25 to 600 (1 to 24") [bar] [PSI]...
  • Page 192: Sensor Jumo Flowtrans Mag H01/02

    13 Technical data 13.3 Sensor JUMO flowTRANS MAG H01/02 13.3.1 Temperatures The temperature range of the device depends on a series of factors. These factors include the measure- ment medium temperature, the ambient temperature, the operating pressure, the lining material, and the approvals for the Ex-protection.
  • Page 193: Maximum Ambient Temperature Depending On The Medium Temperature

    Maximum ambient temperature depending on the medium temperature IMPORTANT (NOTE)! There is the additional document "JUMO flowTRANS MAG S/H - Safety Manual Ex for devices 406012 - 406019" for the measuring systems that are used in potentially explosive areas. The information and data contained in the document must also be strictly observed.
  • Page 194: Signal Cable Length And Preamplifier

    13 Technical data 13.3.10 Signal cable length and preamplifier For devices in the remote mount design, the electrical connection between the transmitter and sensor is made via a signal cable. Transmitter Single-compartment Ex-zone 2 or outside of Ex-area housing housing design Maximum signal cable length Without preamplifier...
  • Page 195: Sensor Materials

    13 Technical data 13.3.11 Sensor materials Component Standard Option touched by me- dium Housing for nominal width: DN 1 to 100 (1/25 to 4") Deep-drawn housing Stainless steel 1.4301 (AISI 304), 1.4308 (ASTM) Terminal box Alum. alloy, painted, (Compact design type) ≥...
  • Page 196: Material Load - Sensor Housing

    13 Technical data 13.3.12 Material load - sensor housing Limits for the admissible fluid temperature (TS) and admissible pressure (PS) are determined by the used liner and flange materials of the device (see device nameplate). Process connection Nominal width (PSI) DN 3 to 50 (1/10"...
  • Page 197 13 Technical data JIS 10K-B2210 Connection flange design Nominal width Material Process connec- Fluid temperature Admissible pressure tion (TS) (PS) DN 32 to 100 1.4404, 1.4435, PN10 -25 to +130 °C 10 bar (145 psi) (1 1/4 to 4") 1.4301 (-13 to +266 °F) Connection flange design...
  • Page 198: Transmitter Jumo Flowtrans Mag 01

    13 Technical data 13.4 Transmitter JUMO flowTRANS MAG 01 13.4.1 Electrical properties Voltage supply AC 100 to 230 V (-15 %/+10 %) AC 24 V (-30 %/+10 %) DC 24 V (-30 %/+30 %), harmonics < 5 % Supply frequency...
  • Page 199: Vibration According To En 60068-2

    13 Technical data 13.4.7 Vibration according to EN 60068-2 Valid for: Remote mount design In the range 10 to 58 Hz (Transmitter) max. 0.15 mm (0.006″) amplitude In the range 58 to 150 Hz max. 2 g acceleration = Peak load...
  • Page 200: Approvals/Approval Marks

    13 Technical data 13.5 Approvals/approval marks JUMO flowTRANS MAG S01/S02 - compact design Model 406012/1-0, 406013/1-0 406012/1-0, 406013/1-0 406012/1-0, 406013/1-0 Terminal box Design type Single-compartment Single-compartment Single-compartment housing housing housing Material Aluminum Aluminum Aluminum Sensor Design type Steel pipe Shell housing...
  • Page 201 13 Technical data JUMO flowTRANS MAG S01/S02 - compact design Model 406012/1-1, 406013/1-1 406012/1-1, 406013/1-1 Terminal box Design type Single-compartment housing Single-compartment housing Material Aluminum Aluminum Sensor Design type Shell housing Shell housing Material Aluminum Aluminum Process connection Material Steel...
  • Page 202 13 Technical data JUMO flowTRANS MAG S01/S02 - compact design Model 406012/1-1, 406013/1-1 406012/1-1, 406013/1-1 Terminal box Design type Dual-compartment housing Dual-compartment housing Material Aluminum Aluminum Sensor Design type Shell housing Shell housing Material Aluminum Aluminum Process connection Material Steel...
  • Page 203 13 Technical data JUMO flowTRANS MAG S01/S02 – remote mount design Model 406012/2-0, 406013/2-0 406012/2-0, 406013/2-0 406012/2-0, 406013/2-0 Terminal box Design type Material Aluminum Aluminum Aluminum Sensor Design type Steel pipe Shell housing Shell housing Material Steel Aluminum Aluminum Process connection...
  • Page 204 13 Technical data JUMO flowTRANS MAG S01/S02 – remote mount design Model 406012/1-1, 406013/1-1 406012/1-1, 406013/1-1 Terminal box Design type Material Aluminum Aluminum Sensor Design type Shell housing Shell housing Material Aluminum Aluminum Process connection Material Steel Stainless steel Ex-protection zone 2...
  • Page 205 13 Technical data JUMO flowTRANS MAG S01/S02 – remote mount design Model 406012/1-1, 406013/1-1 406012/1-1, 406013/1-1 Terminal box Design type Material Aluminum Aluminum Sensor Design type Shell housing Shell housing Material Aluminum Aluminum Process connection Material Steel Stainless steel Ex-protection zone 2...
  • Page 206 13 Technical data JUMO flowTRANS MAG H01/H02 Compact design Remote mount design Model 406015/1-0, 406016/1-0 406015/2-0, 406016/2-0 Terminal box Design type Single-compartment housing Material Aluminum CrNi steel 1.4308 (ASTM) Flange Material CrNi steel 1.4571 (AISI 316 Ti) CrNi steel 1.4571 (AISI 316 Ti)
  • Page 207 13 Technical data JUMO flowTRANS MAG H01/H02 Compact design Remote mount design Model 406015/1-1, 406016/1-1 406015/1-1, 406016/1-1 406015/2-1, 406016/2-1 Terminal box Design type Single-compartment Dual-compartment hous- housing Material Aluminum Aluminum CrNi steel 1.4308 (ASTM) Flange Material CrNi steel 1.4571 CrNi steel 1.4571 CrNi steel 1.4571...
  • Page 208 13 Technical data JUMO flowTRANS MAG 01/02 – remote mount design Model 406018/2-0, 406019/2-0 406018/2-1, 406019/2-1 406018/2-1, 406019/2-1 Terminal box Design type Single-compartment Single-compartment Dual-compartment hous- housing housing Material Aluminum Aluminum Aluminum Ex-protection zone 2 FM15ATEX0024X 2014/34/EU - ATEX FM15ATEX0025X IECEx FME15.0006X...
  • Page 209: 14 Annex

    Iout at Flow > 103 % (No signaling, current output re- mains at 20.5 mA), high alarm, low alarm Low Flow Cut Off 0 to 10 % Empty Pipe Detector ON/OFF TFE Detector ON/OFF Only available for JUMO flowTRANS MAG S02/H02...
  • Page 210: With The Profibus-Pa Version

    Identification marking for intended use in potentially explosive areas according to: • ATEX Directive (identification marking in addition to CE mark) • IECEx standards IMPORTANT (NOTE)! All the documentation, declarations of conformity and certificates are also available in the download area www.jumo.de.
  • Page 211 14 Annex Document No. / Document n°. Manufacturer / Etabli par Address / Adresse Product / Produit Name / Nom Type / Type Data sheet no. / N° Document d'identification We hereby declare in sole responsibility that the designated product fulfills the requirements of the European Directives. Nous déclare sous notre seule responsabilité...
  • Page 212 14 Annex Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Valid for Type / Valable pour le type Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence...
  • Page 213 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 214 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Recognized quality assurance systems of production / Systèmes de qualité reconnus de production Notified Body / Organisme notifié...
  • Page 215 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 216 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 217 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 218 14 Annex Issued by / Etabli par Place, date / Lieu, date Legally binding signatures / Signatures juridiquement valable...
  • Page 219 14 Annex Document No. / Document n°. Manufacturer / Etabli par Address / Adresse Product / Produit Name / Nom Type / Type Data sheet no. / N° Document d'identification We hereby declare in sole responsibility that the designated product fulfills the requirements of the European Directives. Nous déclare sous notre seule responsabilité...
  • Page 220 14 Annex Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Valid for Type / Valable pour le type Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence...
  • Page 221 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 222 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Recognized quality assurance systems of production / Systèmes de qualité reconnus de production Notified Body / Organisme notifié...
  • Page 223 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 224 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 225 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 226 14 Annex Issued by / Etabli par Place, date / Lieu, date Legally binding signatures / Signatures juridiquement valable...
  • Page 227 14 Annex Document No. / Document n°. Manufacturer / Etabli par Address / Adresse Product / Produit Name / Nom Type / Type Data sheet no. / N° Document d'identification We hereby declare in sole responsibility that the designated product fulfills the requirements of the European Directives. Nous déclare sous notre seule responsabilité...
  • Page 228 14 Annex Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Valid for Type / Valable pour le type Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence...
  • Page 229 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 230 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Recognized quality assurance systems of production / Systèmes de qualité reconnus de production Notified Body / Organisme notifié...
  • Page 231 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 232 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 233 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 234 14 Annex Issued by / Etabli par Place, date / Lieu, date Legally binding signatures / Signatures juridiquement valable...
  • Page 235 14 Annex Document No. / Document n°. Manufacturer / Etabli par Address / Adresse Product / Produit Name / Nom Type / Type Data sheet no. / N° Document d'identification We hereby declare in sole responsibility that the designated product fulfills the requirements of the European Directives. Nous déclare sous notre seule responsabilité...
  • Page 236 14 Annex Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Valid for Type / Valable pour le type Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence...
  • Page 237 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 238 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Recognized quality assurance systems of production / Systèmes de qualité reconnus de production Notified Body / Organisme notifié...
  • Page 239 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 240 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 241 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 242 14 Annex Issued by / Etabli par Place, date / Lieu, date Legally binding signatures / Signatures juridiquement valable...
  • Page 243 14 Annex Document No. / Document n°. Manufacturer / Etabli par Address / Adresse Product / Produit Name / Nom Type / Type Data sheet no. / N° Document d'identification We hereby declare in sole responsibility that the designated product fulfills the requirements of the European Directives. Nous déclare sous notre seule responsabilité...
  • Page 244 14 Annex Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Valid for Type / Valable pour le type Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence...
  • Page 245 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 246 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Recognized quality assurance systems of production / Systèmes de qualité reconnus de production Notified Body / Organisme notifié...
  • Page 247 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 248 14 Annex Issued by / Etabli par Place, date / Lieu, date Legally binding signatures / Signatures juridiquement valable...
  • Page 249 14 Annex Document No. / Document n°. Manufacturer / Etabli par Address / Adresse Product / Produit Name / Nom Type / Type Data sheet no. / N° Document d'identification We hereby declare in sole responsibility that the designated product fulfills the requirements of the European Directives. Nous déclare sous notre seule responsabilité...
  • Page 250 14 Annex Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Valid for Type / Valable pour le type Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence...
  • Page 251 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Valid for Type / Valable pour le type EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié...
  • Page 252 14 Annex EU type examination certificate / Certificat d'examen de type UE Reference / Référence Notified Body / Organisme notifié Identification no. / N° d'identification Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque Recognized quality assurance systems of production / Systèmes de qualité reconnus de production Notified Body / Organisme notifié...
  • Page 253 14 Annex Directive / Directive Name / Nom Reference / Référence Comment / Remarque Date of first application of the CE mark to the product / Date de 1ère application du sigle sur le produit Standards/Specifications applied / Normes/Spécifications appliquées Reference / Référence Edition / Édition Comment / Remarque...
  • Page 254 14 Annex Issued by / Etabli par Place, date / Lieu, date Legally binding signatures / Signatures juridiquement valable...
  • Page 255: China Rohs

    14 Annex 14.3 China RoHS...
  • Page 256 14 Annex...
  • Page 257 14 Annex...
  • Page 258 14 Annex...
  • Page 260 JUMO GmbH & Co. KG JUMO Instrument Co. Ltd. JUMO Process Control, Inc. Street address: JUMO House 6733 Myers Road Moritz-Juchheim-Straße 1 Temple Bank, Riverway East Syracuse, NY 13057, USA 36039 Fulda, Germany Harlow, Essex, CM20 2DY, UK Delivery address:...

This manual is also suitable for:

Flowtrans mag s02Flowtrans mag h01Flowtrans mag h02

Table of Contents